Analysis of stress induced by screws in the vertebral fixation system.

Acta Ortop Bras

Universidade de São Paulo, Faculdade de Medicina de Ribeirão Preto, Ribeirão Preto, SP, Brazil, Department de Biomechanics, Medicine and Rehabilitation of the Locomotor System, Faculdade de Medicina de Ribeirão Preto, Universidade de São Paulo, Ribeirão Preto, SP, Brazil.

Published: March 2014

Objective: To compare, using photoelasticity, internal stress produced by USS II type screw with 5.2 and 6.2 mm external diameters, when submitted to three different pullout strengths.

Methods: Two photoelastic models were especially made. The simulation was performed using loads of 1.8, 2.4 e 3.3 kgf.The fringe orders were evaluated around the screws. In all the models analyzed the shear stress were calculated.

Results: Independently of the applied load, the smaller screw showed higher values of shear stress.

Conclusion: According to the analysis performed, we observed that the place of highest stress was in the first thread of the lead, close to the head of the screws. Experimental study.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3952865PMC
http://dx.doi.org/10.1590/S1413-78522014000100002DOI Listing

Publication Analysis

Top Keywords

analysis stress
4
stress induced
4
induced screws
4
screws vertebral
4
vertebral fixation
4
fixation system
4
system objective
4
objective compare
4
compare photoelasticity
4
photoelasticity internal
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!