Coil herniation, premature deployment, and fracture are procedure associated complications of endovascular aneurysm embolization that optimally necessitate coil retrieval when feasible. Several published techniques describe different strategies for managing coil complications including various snare retrieval devices, alligator retrieval devices, stent fixation, and open surgical resection of coils when distal blood flow is compromised. We report a novel technique employed to retrieve a prematurely detached coil during an aneurysm embolization using a syringe fixed to the microcatheter to carefully aspirate a loose coil with direct fluoroscopic visualization. This technique can only be utilized in the circumstance where the proximal end of the coil remains in the microcatheter. Conventional techniques of coil retrieval and stenting are discussed and compared to the rational for using the manual aspiration technique.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3955818PMC
http://dx.doi.org/10.5469/neuroint.2014.9.1.21DOI Listing

Publication Analysis

Top Keywords

aneurysm embolization
12
manual aspiration
8
aspiration technique
8
retrieve prematurely
8
prematurely detached
8
coil
8
detached coil
8
coil retrieval
8
retrieval devices
8
technique
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!