A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Analysis of the effect of renal excretory system cooling during thermal radiofrequency ablation in an animal model. | LitMetric

Analysis of the effect of renal excretory system cooling during thermal radiofrequency ablation in an animal model.

Int Braz J Urol

Faculty of Medicine, Pontifical Catholic University of Campinas, Campinas, SP, Brazil; Radium Institute, Campinas, SP, Brazil; School of Medical Sciences, University of Campinas, Unicamp, Campinas, SP, Brazil.

Published: September 2014

Objective: Analysis of renal excretory system integrity and efficacy of radiofrequency ablation with and without irrigation with saline at 2°C (SF2).

Materials And Methods: The median third of sixteen kidneys were submitted to radiofrequency (exposition of 1 cm) controlled by intra-surgical ultrasound, with eight minutes cycles and median temperature of 90°C in eight female pigs. One excretory renal system was cooled with SF2, at a 30mL/min rate, and the other kidney was not. After 14 days of post-operatory, the biggest diameters of the lesions and the radiological aspects of the excretory system were compared by bilateral ascending pyelogram and the animals were sacrificed in order to perform histological analysis.

Results: There were no significant differences between the diameters of the kidney lesions whether or not exposed to cooling of the excretory system. Median diameter of the cooled kidneys and not cooled kidneys were respectively (in mm): anteroposterior: 11.46 vs. 12.5 (p = 0.23); longitudinal: 17.94 vs. 18.84 (p = 0.62); depth: 11.38 vs. 12.25 (p = 0.47). There was no lesion of the excretory system or signs of leakage of contrast media or hydronephrosis at ascending pyelogram.

Conclusion: Cooling of excretory system during radiofrequency ablation does not sig¬nificantly alter generated coagulation necrosis or affect the integrity of the excretory system in the studied model.

Download full-text PDF

Source
http://dx.doi.org/10.1590/S1677-5538.IBJU.2014.01.14DOI Listing

Publication Analysis

Top Keywords

excretory system
28
radiofrequency ablation
12
analysis renal
8
excretory
8
renal excretory
8
system
8
cooling excretory
8
cooled kidneys
8
system cooling
4
cooling thermal
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!