Purpose: To comparatively evaluate the image quality in normal and forme fruste keratoconus patients using a double-pass system.
Methods: Fourteen normal eyes and 15 eyes with forme fruste keratoconus were included. Corneal tomographic parameters were obtained with optical coherence tomography. Image quality characteristics were determined by an optical quality analysis system (OQAS). Two readings were obtained from each eye. The following parameters were analysed: modulation transfer function cutoff frequency (MTF), Strehl ratio and OQAS values (OV) at contrasts of 100% (OV-100), 20% (OV-20) and 9% (OV-9). Repeatability coefficient (RC), coefficient of variation (CVw) and intraclass correlation coefficient (ICC) were evaluated.
Results: There were no significant differences in central corneal thickness, anterior best-fit sphere and posterior best-fit sphere in both groups. There was a significant difference between the mean MTF (p=0.001), Strehl ratio (p<0.001), OV-100 (p=0.001), OV-20 (p<0.001) and OV-9 (p<0.001) values between both groups. The coefficients of variation were 4.10 (MTF) and 13.77 (Strehl ratio) in normal eyes and 14.53 (MTF) and 14.97 (Strehl ratio) in eyes with forme fruste keratoconus. The ICC values were 0.99 and 0.95 for MTF and, 0.80 and 0.91 for Strehl ratio, in normal and forme fruste keratoconus eyes, respectively. The corresponding repeatability values were 3.53 and 6.15 for MTF and, 0.06 and 0.04 for Strehl ratio, respectively.
Conclusions: OQAS has the potential to be a useful diagnostic tool in cases with early keratoconus. The repeatability of a double-pass (DP) system was good in normal as well as forme fruste keratoconus eyes.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1136/bjophthalmol-2013-304494 | DOI Listing |
J Cardiothorac Vasc Anesth
January 2025
Division of Cardiovascular & Thoracic Anesthesia, Department of Anesthesiology, Perioperative and Pain Medicine, Stanford University School of Medicine, Stanford, CA.
J Curr Ophthalmol
October 2024
Department of Ophthalmology, Rayong Hospital, Rayong, Thailand.
Cornea
November 2024
Department of Ophthalmology and Visual Sciences, Paulista School of Medicine, Federal University of São Paulo - UNIFESP, São Paulo, Brazil; and.
Purpose: To assess the accuracy of various Pentacam indices in distinguishing keratoconus (KC) in pediatric patients with high astigmatism and to establish appropriate cutoff values.
Methods: This prospective multicenter cross-sectional study included 312 eyes from 167 patients aged 6 to 18 years (mean age, 13.1 ± 3.
Indian J Pediatr
January 2025
Department of Pediatrics, Advanced Pediatrics Centre, Postgraduate Institute of Medical Education and Research, Chandigarh, 160012, India.
BMC Ophthalmol
October 2024
Eye Institute and Department of Ophthalmology, Eye & ENT Hospital, Fudan University, Shanghai, 200031, China.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!