Purpose: Contact with cardiac tissue is a determinant of lesion efficacy during atrial fibrillation (AF) ablation. The Sensei®X Robotic Catheter System (Hansen Medical, CA) has been validated for contact force sensing. The electrical coupling index (ECI) from the EnSite Contact™ system (St. Jude Medical, MN) has been validated as an indicator of tissue contact. We aimed at analyzing ECI behavior during radiofrequency (RF) pulses maintaining a stable contact through the robotic navigation contact system.
Methods: In 15 patients (age, 59 ± 12) undergoing AF ablation, pulmonary vein (PV) isolation was guided by the Sensei®X System, employing the Contact™ catheter.
Results: During the procedure, we assessed ECI changes associated with adequate contact based on the IntelliSense® force-sensing technology (Hansen Medical, CA. Baseline contact (27 ± 8 g/cm(2)) ECI value was 99 ± 13, whereas ECI values in a noncontact site (0 g/cm(2)) and in a light contact site (1-10 g/cm(2)) were respectively 66 ± 12 and 77 ± 10 (p < 0.0001). Baseline contact ECI values were not different depending on AF presentation (paroxysmal AF, 98 ± 9; persistent AF, 100 ± 9) or on cardiac rhythm (sinus rhythm, 97 ± 7; AF,101 ± 10). In all PVs, ECI was significantly reduced during and after ablation (ECI during RF, 56 ± 15; ECI after RF, 72 ± 16; p < 0.001). A mean reduction of 32.2% during RF delivery and 25.4% immediately after RF discontinuation compared with baseline ECI was observed.
Conclusions: Successful PV isolation is associated with a significant decrease in ECI of at least 20 %. This may be used as a surrogate marker of effective lesion in AF ablation.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s10840-014-9882-2 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!