Three-dimensional and chemical changes on the surface of a 3-year clinically retrieved oxidized titanium dental implant.

J Mech Behav Biomed Mater

Department of Oral and Maxillofacial-Head and Neck Surgery, University Hospital Infanta Cristina, Badajoz, Spain; FEDICOM, Foundation for the Study and Development in Implantology, Oral and Maxillofacial Surgery, Badajoz, Spain; CICOM, Centre for Implantology, Oral and Maxillofacial Surgery, Badajoz, Spain.

Published: June 2014

Objective: To report the main topographical features in the micro- and nano-scales and to assess implant chemical changes of the surface of a 3-year clinically retrieved oxidized titanium dental implant, and compare them with a similar, unused implant.

Materials And Methods: The surface of the oxidized titanium dental implants was assessed by surface electron microscopy (SEM) analysis at increasing magnifications. X-ray photoelectron spectroscopy (XPS) measurement was performed to analyze the implants surface chemistry. XPS spectra were acquired before and after sputtering with an Ar(+) ion etching of 3keV.

Results: With a length of 10-40μm, and a width of 0.05-0.1μm, numerous cracks were ubiquitous along the implant surfaces. Chimney-like structures formed micropores between 1 and 5μm, with up to 40% of them partially or totally broken in the retrieved implant. In relation to chemical composition, Ti and O were predominant in both the unused and in the retrieved implant. N was present in high concentrations (11.49at%) at the retrieved implant surface, in contrast with those observed for the unused implant (1.14at%). Also, C was present in higher concentrations in the retrieved implant surface, while drastically decreased following the sputter-cleaning process.

Conclusion: While cracks were ubiquitous present from the manufacturing, broken chimney-like structures forming micropores in the clinically retrieved implant may be attributable to excessive mechanical friction forces during the insertion of the implant. Chemical composition of the implant surface may be subjected to changes because of the in vivo environment, with increase of N and C, and decrease of Ca and P.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jmbbm.2014.02.020DOI Listing

Publication Analysis

Top Keywords

retrieved implant
20
clinically retrieved
12
oxidized titanium
12
titanium dental
12
implant
12
implant surface
12
chemical changes
8
surface
8
changes surface
8
surface 3-year
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!