In silico study of low-frequency transcranial ultrasound fields in acute ischemic stroke patients.

Ultrasound Med Biol

Division of Cardiovascular Health and Disease, Department of Internal Medicine, College of Medicine, University of Cincinnati, Cincinnati, OH, USA; Biomedical Engineering Program, College of Engineering and Applied Sciences, University of Cincinnati, Cincinnati, OH, USA. Electronic address:

Published: June 2014

Ultrasound in the sub-megahertz range enhances thrombolysis and may be applied transcranially to ischemic stroke patients. The consistency of transcranial insonification needs to be evaluated. Acoustic and thermal simulations based on computed-tomography (CT) scans of 20 patients were performed. An unfocused 120-kHz transducer allowed homogeneous insonification of the thrombus, and positioning based on external landmarks performed similarly to an optimized placement based on CT data. With a weakly focused 500-kHz transducer, the landmark-based positioning underperformed. The predicted inter-patient variation of in situ acoustic pressure was similar with both the 120 and 500-kHz transducers for the optimized placement (18.0-26.4% relative standard deviation). The simulated maximum acoustic pressure in intervening tissues was 2.6 ± 0.6 and 2.0 ± 0.7 times the pressure in the thrombus for the 120-kHz and 500-kHz transducers, respectively. A 1 W/cm(2) insonification of the thrombus caused a 3.8 ± 2.2 °C increase in the bone for the 120-kHz transducer, and a 13.4 ± 3.3 °C increase for the 500-kHz transducer. Contralateral local maxima up to 1.1 times the pressure amplitude in the targeted zone were predicted for the 120-kHz transducer. We established two transducer placement approaches, one based on analysis of a head CT and the other using simple external, visible landmarks. Both approaches allowed consistent insonification of the thrombus.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4012005PMC
http://dx.doi.org/10.1016/j.ultrasmedbio.2013.12.025DOI Listing

Publication Analysis

Top Keywords

120-khz transducer
12
insonification thrombus
12
ischemic stroke
8
stroke patients
8
optimized placement
8
500-khz transducer
8
acoustic pressure
8
500-khz transducers
8
times pressure
8
°c increase
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!