Lipemia: causes, interference mechanisms, detection and management.

Biochem Med (Zagreb)

University Department of Chemistry, Medical School University Hospital Sestre Milosrdnice, Zagreb, Croatia.

Published: April 2014

In the clinical laboratory setting, interferences can be a significant source of laboratory errors with potential to cause serious harm for the patient. After hemolysis, lipemia is the most frequent endogenous interference that can influence results of various laboratory methods by several mechanisms. The most common preanalytical cause of lipemic samples is inadequate time of blood sampling after the meal or parenteral administration of synthetic lipid emulsions. Although the best way of detecting the degree of lipemia is measuring lipemic index on analytical platforms, laboratory experts should be aware of its problems, like false positive results and lack of standardization between manufacturers. Unlike for other interferences, lipemia can be removed and measurement can be done in a clear sample. However, a protocol for removing lipids from the sample has to be chosen carefully, since it is dependent on the analytes that have to be determined. Investigation of lipemia interference is an obligation of manufacturers of laboratory reagents; however, several literature findings report lack of verification of the declared data. Moreover, the acceptance criteria currently used by the most manufacturers are not based on biological variation and need to be revised. Written procedures for detection of lipemia, removing lipemia interference and reporting results from lipemic samples should be available to laboratory staff in order to standardize the procedure, reduce errors and increase patient safety.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3936974PMC
http://dx.doi.org/10.11613/BM.2014.008DOI Listing

Publication Analysis

Top Keywords

lipemia interference
12
lipemic samples
8
lipemia
7
laboratory
6
interference mechanisms
4
mechanisms detection
4
detection management
4
management clinical
4
clinical laboratory
4
laboratory setting
4

Similar Publications

Background: Grossly lipaemic samples are a significant cause of analytical errors, potentially impacting patient care. The causes of lipaemia are varied and often unavoidable, while methods to reduce lipaemia through gold-standard ultracentrifugation are limited by availability, transportation and cost. Benchtop centrifugation has been proposed as an alternative method to reduce lipaemia.

View Article and Find Full Text PDF

Understanding green discoloration in serum and citrate plasma samples: a case report.

Biochem Med (Zagreb)

February 2025

Department of Medical Biochemistry and Laboratory Medicine, Merkur University Hospital, Zagreb, Croatia.

This case report investigates the occurrence of green discoloration in serum and citrate plasma samples collected from a male adult patient following a multivisceral organ transplant. In collected samples, it was necessary to investigate the influence of sample discoloration on the results of laboratory tests and to determine the appropriate approach to sample management. Hematology, coagulation and blood gas analysis showed no flags, but the biochemical lipemia index was susceptible to positive interference, necessitating dilution of the native sample.

View Article and Find Full Text PDF
Article Synopsis
  • * A case study of a severely burned patient showed triglyceride levels reaching 5,696 mg/dL, which did not match the clear plasma observed and low serum apolipoprotein B levels, indicating an issue with measurement rather than an actual health problem.
  • * The falsely elevated triglyceride and lipase levels were attributed to high serum glycerol levels caused by the use of silver sulfadiazine cream on burn wounds, which affected lab results; levels normalized once the cream
View Article and Find Full Text PDF

Biochemical exploration of cholestasis: interpretation, traps and interferences.

Clin Biochem

January 2025

Biochemistry Laboratory, CHU Pontchaillou, Rennes, France; Univ Rennes, CHU Rennes, INSERM, EHESP, IRSET (Institut de Recherche en Santé, Environnement et Travail) UMR_S 1085, F-35000 Rennes, France. Electronic address:

We described the case of a 33-year-old patient who presented to the emergency department with non-febrile jaundice associated with epigastric pain. He suffered from acute non-severe alcoholic hepatitis and cholestasis. Biochemical investigations highlighted a huge elevation of the alpha-1-globulins fraction with an unexpected peak in the alpha-1-globulins area in serum protein electrophoresis, a severe hypercholesterolemia without xanthelasmas nor cholesterolomas.

View Article and Find Full Text PDF

The erythrocyte sedimentation rate (ESR) is a widely used diagnostic test, influenced by all physiological and pathological conditions that can bias blood rheology by interfering factors. This study aimed to evaluate the performance of the CUBE 30 touch ESR analyzer in samples with preanalytical variables, as lipemia, hemolysis, and icterus or in presence of fibrinogen., Moreover we focused to define the maximum time limits to ensure a reliable ESR measure.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!