Validity Test of a New Open-Circuit Indirect Calorimeter.

JPEN J Parenter Enteral Nutr

Department of Clinical Nutrition, Department of Nursing, Penn State Milton S. Hershey Medical Center, Hershey, Pennsylvania.

Published: August 2015

Background: Indirect calorimetry is an accurate way to measure resting metabolic rate. The Deltatrac Metabolic Monitor is considered a criterion standard but is no longer manufactured. New-generation indirect calorimeters have been introduced, but there are limited published validation data comparing these devices to criterion instruments.

Materials And Methods: A prospective, observational, N-of-1 trial was conducted to validate a new-generation indirect calorimeter against a gold standard device. This design was chosen to minimize and define the degree of biological variation, thus focusing on variation due to the devices. Measurements of gas exchange using both indirect calorimeters were conducted daily for 10 consecutive days. Another set of measurement pairs was conducted using just the criterion device for 10 days. Ninety-five percent confidence intervals of differences were used to test for bias. Precision was defined as repeat measures with one device falling within 5% of the other at least 90% of the time.

Results: There were no statistically significant differences between the devices for any measured or calculated parameter. Interdevice differences were no larger than intradevice differences using the criterion instrument. The values obtained from the new device were precise and unbiased compared with the values obtained from the gold standard device.

Conclusion: The new indirect calorimeter measures gas exchange in a reliable and accurate manner compared with a gold standard device. The two devices are equivalent.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4160429PMC
http://dx.doi.org/10.1177/0148607114526242DOI Listing

Publication Analysis

Top Keywords

indirect calorimeter
12
gold standard
12
new-generation indirect
8
indirect calorimeters
8
standard device
8
gas exchange
8
indirect
6
device
5
validity test
4
test open-circuit
4

Similar Publications

Background: Sixty minutes is currently the shortest testing duration for 24-h resting energy expenditure (24-h REE) utilizing whole-room indirect calorimetry.

Objective: Show that recalculated 30-min extrapolated 24-h REE from previously published 60-min metabolic data are valid.

Methods: Propane consumption linearity was determined through an 8-h combustion test.

View Article and Find Full Text PDF

The usefulness of a new indirect calorimeter in critically ill adult patients.

Clin Nutr

October 2024

Department of Neonatal and Pediatric Intensive Care, Division of Pediatric Intensive Care, Erasmus Medical Centre - Sophia Children's Hospital, Rotterdam, the Netherlands.

Background & Aims: The use of indirect calorimetry to determine energy requirements is highly recommended in critically ill patients. To facilitate this a new and easy to use calorimeter (Q-NRG+, Cosmed) was developed. The primary aim of our study was to describe the usefulness of this calorimeter and, secondarily, to investigate the agreement between measured and predicted energy needs in a large cohort of critically ill adult patients.

View Article and Find Full Text PDF

Alga 1.0, a product containing bromoform, was fed to cattle to evaluate its effects on methane ( ) and carbon dioxide ( ) emissions and diet digestibility. Twelve nonlactating, nonpregnant Jersey cows (490 ± 19 kg body weight) were used in four replicated 3 × 3 Latin squares with three periods, each consisting of 21 d.

View Article and Find Full Text PDF

Dietary macronutrient composition and its effect on 24-h substrate oxidation: A study of diurnal variations in carbohydrate and fat intake.

Clin Nutr

September 2024

Department of Nutrition and Metabolism, National Institute of Health and Nutrition, National Institutes of Biomedical Innovation, Health and Nutrition, Settsu, Osaka 566-0002, Japan; Faculty of Nutrition, Kagawa Nutrition University, Sakado, Saitama 350-0288, Japan; Institute of Nutrition Sciences, Kagawa Nutrition University, Sakado, Saitama 350-0288, Japan. Electronic address:

Background & Aims: In recent times, the complexity of food styles and meal content has increased, leading to significant variations in macronutrient composition between meals. This phenomenon has coincided with a rise in obesity rates. We aimed to determine whether a large variation in macronutrient composition between meals results in reduced fat oxidation.

View Article and Find Full Text PDF

Evaluating predictive equations for energy requirements throughout breast cancer trajectory: A comparative study.

Clin Nutr

September 2024

Human Nutrition Research Unit, Department of Agricultural, Food & Nutritional Science, University of Alberta, Edmonton, Alberta, Canada. Electronic address:

Article Synopsis
  • The study aimed to evaluate the accuracy of 40 predictive equations for estimating resting energy expenditure (REE) in women with breast cancer, comparing these predictions to measurements taken by indirect calorimetry.
  • The analysis included data from 90 participants (24% survivors, 61.1% early-stage cancer), revealing that none of the equations provided accurate REE estimates both at the group and individual levels, with significant biases observed.
  • These findings highlight the importance of using precise methods for calculating energy needs in women with breast cancer, as inaccuracies can impact treatment and recovery by influencing energy intake.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!