In-beam positron emission tomography (PET) can enable visualization of an irradiated field using positron emitters (β+ decay). In particle therapies, many kinds of secondary particles are produced by nuclear interactions, which affect PET imaging. Our purpose in this work was to evaluate effects of secondary particles on in-beam PET imaging using the Monte Carlo simulation code, Geant4, by reproducing an experiment with a small OpenPET prototype in which a PMMA phantom was irradiated by a (11)C beam. The number of incident particles to the detectors and their spectra, background coincidence for the PET scan, and reconstructed images were evaluated for three periods, spill-time (beam irradiation), pause-time (accelerating the particles) and beam-off time (duration after the final spill). For spill-time, we tested a background reduction technique in which coincidence events correlated with the accelerator radiofrequency were discarded (RF gated) that has been proposed in the literature. Also, background generation processes were identified. For spill-time, most background coincidences were caused by prompt gamma rays, and only 1.4% of the total coincidences generated β+ signals. Differently, for pause-time and beam-off time, more than 75% of the total coincidence events were signals. Using these coincidence events, we failed to reconstruct images during the spill-time, but we obtained successful reconstructions for the pause-time and beam-off time, which was consistent with the experimental results. From the simulation, we found that the absence of materials in the beam line and using the RF gated technique improved the signal-to-noise ratio for the spill-time. From an additional simulation with range shifter-less irradiation and the RF gated technique, we showed the feasibility of image reconstruction during the spill-time.

Download full-text PDF

Source
http://dx.doi.org/10.1088/0031-9155/59/7/1623DOI Listing

Publication Analysis

Top Keywords

secondary particles
12
beam-off time
12
coincidence events
12
monte carlo
8
carlo simulation
8
small openpet
8
openpet prototype
8
11c beam
8
beam irradiation
8
effects secondary
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!