Objective: This study was done to assess breast density on digital mammography and digital breast tomosynthesis according to the visual Breast Imaging Reporting and Data System (BI-RADS) classification, to compare visual assessment with Quantra software for automated density measurement, and to establish the role of the software in clinical practice.
Materials And Methods: We analysed 200 digital mammograms performed in 2D and 3D modality, 100 of which positive for breast cancer and 100 negative. Radiological density was assessed with the BI-RADS classification; a Quantra density cut-off value was sought on the 2D images only to discriminate between BI-RADS categories 1-2 and BI-RADS 3-4. Breast density was correlated with age, use of hormone therapy, and increased risk of disease.
Results: The agreement between the 2D and 3D assessments of BI-RADS density was high (K 0.96). A cut-off value of 21% is that which allows us to best discriminate between BI-RADS categories 1-2 and 3-4. Breast density was negatively correlated to age (r = -0.44) and positively to use of hormone therapy (p = 0.0004). Quantra density was higher in breasts with cancer than in healthy breasts.
Conclusions: There is no clear difference between the visual assessments of density on 2D and 3D images. Use of the automated system requires the adoption of a cut-off value (set at 21%) to effectively discriminate BI-RADS 1-2 and 3-4, and could be useful in clinical practice.