Introduction: This report presents safety data on the use of a new open-irrigation radiofrequency ablation (RFA) catheter with a 56-hole porous tip in 742 patients enrolled in 2 US prospective, multicenter observational registry studies representing real-world use of the catheter.

Methods: This analysis is comprised of patients who underwent RFA of drug-refractory recurrent symptomatic paroxysmal atrial fibrillation (AF). Acute adverse events (AEs) were collected and categorized by seriousness, timing, and relatedness, with 7 days of follow-up data in one study and at least 120 days of data from a 1-year follow-up in the other. Acute serious adverse events (SAEs) that were identified as potentially related to the device and/or procedure were adjudicated by an independent safety committee.

Results: A total of 30 patients (4.0%) in the combined studies experienced an acute SAE related to the device and/or procedure, which was similar in the subset of patients age 65 and over (4.2%). These SAEs included 1.2% cardiac tamponade/perforation, 0.7% pericarditis, 0.5% pulmonary events, and 0.8% vascular access complications. No myocardial infarction, stroke, transient ischemic attack, or atrioesophageal fistulas within 7 days postprocedure were reported. In the study with extended follow-up, 1 pulmonary vein stenosis and 1 esophageal injury were seen beyond 7 days postprocedure (0.2% each). There were no device or procedure related deaths.

Conclusion: Results from 2 large observational studies demonstrated that a new porous tip RFA catheter was safe for the treatment of drug refractory, recurrent, symptomatic paroxysmal AF, including treatment of older patients (≥65 years).

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4282391PMC
http://dx.doi.org/10.1111/jce.12403DOI Listing

Publication Analysis

Top Keywords

catheter 56-hole
8
56-hole porous
8
radiofrequency ablation
8
paroxysmal atrial
8
atrial fibrillation
8
observational registry
8
registry studies
8
rfa catheter
8
recurrent symptomatic
8
symptomatic paroxysmal
8

Similar Publications

Introduction: A 56-hole porous tip radiofrequency catheter was developed to provide more uniform cooling with less fluid delivery than a prior 6-hole irrigated design. This study aimed to evaluate the impact of contact force (CF) ablation with the porous tip on complications (congestive heart failure [CHF] and non-CHF related), healthcare resource utilization, and procedural efficiency in patients undergoing de novo paroxysmal atrial fibrillation (PAF) ablations in a real-world setting.

Methods: Consecutive de novo PAF ablations were performed between February 2014 and March 2019 by six operators at a single US academic center.

View Article and Find Full Text PDF

Importance: The ThermoCool SmartTouch catheter (ablation catheter with contact force and 6-hole irrigation [CF-I6]) is approved by the US Food and Drug Administration (FDA) for paroxysmal atrial fibrillation (AF) ablation and used in routine clinical practice for persistent AF ablation, although clinical outcomes for this indication are unknown. There is a need to understand whether data from routine clinical practice can be used to conduct regulatory-grade evaluations and support label expansions.

Objective: To use health system data to compare the safety and effectiveness of the CF-I6 catheter for persistent AF ablation with the ThermoCool SmartTouch SurroundFlow catheter (ablation catheter with contact force and 56-hole irrigation [CF-I56]), which is approved by the FDA for this indication.

View Article and Find Full Text PDF

Purpose: The prospective, multicenter SMART SF trial demonstrated the acute safety and effectiveness of the 56-hole porous tip irrigated contact force (CF) catheter for drug-refractory paroxysmal atrial fibrillation (PAF) ablation with a low primary adverse event rate (2.5%), leading to FDA approval of the catheter. Here, we are reporting the long-term effectiveness and safety results that have not yet been reported.

View Article and Find Full Text PDF

Contact force sensing in ablation of ventricular arrhythmias using a 56-hole open-irrigation catheter: a propensity-matched analysis.

J Interv Card Electrophysiol

April 2021

Cardiology Clinical Academic Group, St. George's University Hospitals NHS Foundation Trust, St. George's, University of London, London, UK.

Purpose: The effect of adding contact force (CF) sensing to 56-hole tip irrigation in ventricular arrhythmia (VA) ablation has not been previously studied. We aimed to compare outcomes with and without CF sensing in VA ablation using a 56-hole radiofrequency (RF) catheter.

Methods: A total of 164 patients who underwent first-time VA ablation using Thermocool SmartTouch Surround Flow (TC-STSF) catheter (Biosense-Webster, Diamond Bar, CA, USA) were propensity-matched in a 1:1 fashion to 164 patients who had first-time ablation using Thermocool Surround Flow (TC-SF) catheter.

View Article and Find Full Text PDF

Aims: This study aimed to evaluate feasibility and safety as well as 1-year clinical outcome of pulmonary vein isolation (PVI) using a unique radiofrequency ablation catheter ("Thermocool SmartTouch SurroundFlow"; STSF) incorporating both, contact force (CF) sensing technology and enhanced tip irrigation with 56 holes, in one device.

Methods: A total of 110 patients suffering from drug-refractory atrial fibrillation underwent wide area circumferential PVI using either the STSF ablation catheter (75 consecutive patients, study group) or a CF catheter with conventional tip irrigation ("Thermocool SmartTouch", 35 consecutive patients, control group). For each ablation lesion, a target CF of ≥ 10-39 g and a force time integral (FTI) of > 400 g s was targeted.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!