Background: Intensity modulated radiation therapy (IMRT) dosimetry verification is routinely conducted via integrated or individual field dosimetry using film or a matrix of detectors. Techniques and software systems are commercially available which use individual field dosimetry measurements as input into algorithms that estimate 3D patient dose distributions on CT scan derived target volumes and organs at risk (OARs), thus allowing direct dose-volume histogram (DVH) analysis vs. treatment planning system (TPS) DVH. The purpose of this work is to present a systematic benchmarking technique to evaluate the accuracy and consistency of such a software system.

Methods: A MapCheck2 diode array and 3DVH™ software from Sun Nuclear were used for this study. Delivered planar dose was measured with the diode array as an input to 3DVH™ software that was used to estimate the 3D dose matrix. Accuracy of the output of 3DVH™ is tested by comparing measured planar doses over a range of depths to the same planes reconstructed by 3DVH™. Different fields from complex IMRT cases were selected and examined in this study. The sensitivity to depth of measurement was evaluated.

Results: The Gamma Index analysis, comparing calculated 3D dose with measured 3D dose with 2% and 2mm distance-to-agreement (DTA) criteria returned a pass rate of > 90% for all patient cases calculated by the treatment planning system and it returned a pass rate of > 96% in 9 out of 10 cases calculated by 3DVH™. Extracted computed dose planes with 3DVH™ software at different depths in the flat phantom passed all gamma evaluation analyses when compared to measured planes at different depths using MapCheck2.

Conclusions: Studying complex head and neck IMRT fields, it was shown that the 3D dose distribution predicted by the planned dose perturbation (PDP) algorithm is both accurate and consistent.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3908853PMC
http://dx.doi.org/10.2478/raon-2013-0042DOI Listing

Publication Analysis

Top Keywords

3dvh™ software
12
dose
9
planned dose
8
individual field
8
field dosimetry
8
treatment planning
8
planning system
8
diode array
8
dose measured
8
returned pass
8

Similar Publications

Thanks to the plummeting costs of continuously evolving omics analytical platforms, research centers collect multiomics data more routinely. They are, however, confronted with the lack of a versatile software solution to harmoniously analyze single-omics and interpret multiomics data. We have developed iSODA, a web-based application for the analysis of single- and multiomics data.

View Article and Find Full Text PDF

Background Artificial Intelligence (AI) is revolutionizing medical science, with significant implications for radiology. Understanding the knowledge, attitudes, perspectives, and practices of medical professionals and residents related to AI's role in radiology is crucial for effective integration. Methods A cross-sectional survey was conducted among members of the Indian Radiology & Imaging Association (IRIA), targeting practicing radiologists and residents across academic and non-academic institutions.

View Article and Find Full Text PDF

Hyperspectral images are rich in spectral and spatial information, providing a detailed and comprehensive description of objects, which makes hyperspectral image analysis technology essential in intelligent agriculture. With various corn seed varieties exhibiting significant internal structural differences, accurate classification is crucial for planting, monitoring, and consumption. However, due to the large volume and complex features of hyperspectral corn image data, existing methods often fall short in feature extraction and utilization, leading to low classification accuracy.

View Article and Find Full Text PDF

Purpose//objectives: A disproportionate incidence's increase of rectal cancer in patients younger than 50 years of age. The ESMO and NCCN recommendations are not age-specific and the literature is poor and conflicting. We decided to examine patients with rectal cancer treated in our centre in the last 15 years with curative neoadjuvant radiochemotherapy comparing outcomes in the two groups under and over 55 years old.

View Article and Find Full Text PDF

Post-comatose disorders of consciousness (DoC) represent persistent neurological conditions with limited therapeutic options and a poor prognosis. Recent works advocate for exploring the effects of psychedelics to enhance brain complexity in DoC and ameliorate their consciousness. We investigated sub-anesthetic concentration of the atypical psychedelic ketamine for treating post-comatose prolonged DoC through a double-blind, placebo-controlled, cross-over trial involving three adult patients.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!