A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 143

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 143
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 209
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 980
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3077
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 574
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 488
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Fluorescence immunoassay of human D-dimer in whole blood. | LitMetric

Fluorescence immunoassay of human D-dimer in whole blood.

J Clin Lab Anal

Department of Biomedical Science, Hallym University, Chuncheon, South Korea; BodiTech Med, Inc, Chuncheon, South Korea.

Published: July 2014

Background: D-dimer is a widely used biomarker for the initial clinical assessment of suspected deep vein thrombosis and pulmonary embolism. Here, we presented a new fluorescence (FL) D-dimer assay system, which was developed with a platform of point-of-care test (POCT) for clinical applications.

Methods: Whole blood was mixed with FL-labeled anti-D-dimer detector antibody, and then loaded onto a disposable cartridge. After 12 min of incubation, the FL intensity was acquired by scanning of test cartridge and converted as level of D-dimer in a laser FL scanner. The analytical performance of FL immunoassay was evaluated by linearity, recovery, and precision tests. The comparability of the developed assay was examined with automated reference methods.

Results: The FL assay system showed a good linearity, and the analytical mean recovery of control was 103% in a dynamic working range. The imprecision of intra- and inter-as-say of coefficient of variations from assay system was less than 8%. The developed FL assay system showed strong correlation with two automated reference assays, Vidas D-dimer (r = 0.973) and Stalia D-dimer (r = 0.971).

Conclusion: The new FL immunoassay for D-dimer is a user-friendly, precise, and reproducible platform of POCT in whole blood.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6807400PMC
http://dx.doi.org/10.1002/jcla.21683DOI Listing

Publication Analysis

Top Keywords

assay system
16
system developed
8
developed assay
8
automated reference
8
d-dimer
7
assay
5
fluorescence immunoassay
4
immunoassay human
4
human d-dimer
4
d-dimer blood
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!