Purpose. To evaluate the long-term efficacy, accuracy, stability, and safety of hyperopic laser in situ keratomileusis (LASIK) using a 213 nm wavelength solid-state laser. Methods. This prospective noncomparative case series consisted of 34 eyes of 17 patients which underwent hyperopic LASIK using a 213 nm solid-state laser (Pulzar Z1, CustomVis) at an outpatient refractive surgery center in Manila, Philippines. The preoperative and postoperative examinations included uncorrected distance visual acuity (UDVA), subjective manifest refraction, corrected distance visual acuity (CDVA), cycloplegic refraction, slitlamp biomicroscopy, and keratometry (K). Main Outcome Measures. Accuracy, efficacy, stability, and safety of the refractive procedure. Results. Mean follow-up was 25.18 ± 13.79 months. At the end of follow-up, 26.47% had a UDVA of 20/20 and 94.12% had a UDVA of ≥20/40. Manifest refractive spherical equivalent (MRSE) was within ±0.50 D of the target refraction in 55.88% and within ±1.0 D in 85.30% of the study eyes. Refractive stability was noted in the 1st postoperative month while hyperopic regression was noted after the 3rd postoperative year. No eye lost more than 2 lines of CDVA. Conclusion. Our results show that the 213 nm solid state laser system is safe, effective, accurate, and predictable for the treatment of hyperopia.