Comparison of residual shunt rates in five devices used to treat patent foramen ovale.

Catheter Cardiovasc Interv

Program in Interventional Cardiology Division of Cardiology, David Geffen School of Medicine, University of California at Los Angeles, Los Angeles, California.

Published: September 2014

Objectives: To assess the effective closure rate among devices used for transcatheter patent foramen ovale (PFO) closure, and to discuss the management of patients with large residual shunts.

Background: Several devices are used off-label for transcatheter closure of a PFO in the United States. The rate of residual shunting after PFO closure varies by device. Failure of effective closure poses risk of a recurrent cerebrovascular event, persistent migraine, or recurrent orthodeoxia.

Methods: Patients who underwent PFO closure in the Cardiac Catheterization Laboratory at UCLA between 2001 and 2013 and had baseline and adequate follow-up transcranial Doppler studies following device placement were enrolled in the study.

Results: Of 167 patients whose records were analyzed, effective PFO closure occurred in 90% (150/167) of patients. The highest effective closure rate was with the Amplatzer Septal Occluder (ASO; 100%), followed by the Amplatzer Cribriform (93%), Gore Helex (90%), Amplatzer PFO (86%), and CardioSEAL (86%) device. The highest rate of residual shunting was observed after placement of the 30-mm Gore Helex device (55%). Of the 17 patients with a residual shunt, three required a repeat PFO closure procedure due to a significant residual shunt associated with recurrent pulmonary emboli or profound orthodeoxia. All three patients received an ASO which successfully closed the residual shunt.

Conclusions: Transcatheter PFO closure has a high success rate, but a moderate residual shunt occurs in about 10% of cases. The observed incidence of residual shunting after PFO closure is significantly larger with the 30-mm Helex device. © 2014 Wiley Periodicals, Inc.

Download full-text PDF

Source
http://dx.doi.org/10.1002/ccd.25453DOI Listing

Publication Analysis

Top Keywords

pfo closure
28
residual shunt
16
effective closure
12
residual shunting
12
closure
11
pfo
9
patent foramen
8
foramen ovale
8
closure rate
8
residual
8

Similar Publications

Effects of patent foramen ovale in migraine: a metabolomics-based study.

J Physiol

January 2025

Department of Neurology, West China Hospital, Sichuan University, Chengdu, China.

Patent foramen ovale (PFO), a cardiac anatomical anomaly inducing abnormal haemodynamics, leads to a paradoxical bypass of the pulmonary circulation. PFO closure might alleviate migraines; however, clinical evidence and basic experiments for the relationship are lacking. To explore the effect of PFO on migraine, 371 migraineurs finishing blood tests and contrast transthoracic echocardiography for the detection of PFO were prospectively included.

View Article and Find Full Text PDF

Background: Since patients with congenital heart defects (CHD) frequently require life-long medical care and repeat invasive treatment, radiation exposure during interventional procedures is a relevant issue concerning potential radiation related risks. Therefore, an analysis on radiation data from the German Registry for Cardiac Operations and Interventions in patients with CHD was performed.

Methods: From January 2012 until December 2020 a total of 28,374 cardiac catheter interventions were recorded.

View Article and Find Full Text PDF

This study aimed to investigate the metabolic mechanisms underlying the combination of patent foramen ovale (PFO) and migraine by assessing metabolite expression before and after interventional occlusion surgery. The study included 11 PFO patients from the Heart Center of Xinjiang Medical University Affiliated Hospital of Traditional Chinese Medicine, who underwent transcatheter PFO intervention and occlusion surgery between January 2018 and February 2023, and 11 healthy controls. Blood samples were collected pre-surgery, 3 days post-surgery, and 30 days post-surgery for metabolomics analysis.

View Article and Find Full Text PDF

We aimed to evaluate the incidence of residual shunt after patent foramen ovale (PFO) closure and analyze the anatomical features of PFO to determine the risk factors for significant residual shunt after PFO closure. Ninety-two patients who underwent PFO closure at our center between September 2021 and June 2022 were consecutively enrolled. Transthoracic saline contrast echocardiography was performed at 6 and 12 months postoperatively to evaluate the presence of a significant residual shunt.

View Article and Find Full Text PDF

Arrhythmias Following Patent Foramen Ovale Closure: An Unsolved Enigma.

Life (Basel)

December 2024

First Department of Cardiology, Hippokration Hospital, Athens Medical School, National and Kapodistrian University of Athens, Vas. Sofias 114, 11527 Athens, Greece.

Patent foramen ovale (PFO) closure has proven to be an effective method of reducing the risk of recurrent stroke in patients with embolic stroke of unknown origin (ESUS). One of the most recognized post-procedural complications is the de novo occurrence of supraventricular arrhythmias, mainly atrial fibrillation, in the first three months following PFO closure. Earlier studies reported the incidence to be around 3.