Purpose: To develop and evaluate fully automated volumetric modulated arc therapy (VMAT) treatment planning for prostate cancer patients, avoiding manual trial-and-error tweaking of plan parameters by dosimetrists.

Methods And Materials: A system was developed for fully automated generation of VMAT plans with our commercial clinical treatment planning system (TPS), linked to the in-house developed Erasmus-iCycle multicriterial optimizer for preoptimization. For 30 randomly selected patients, automatically generated VMAT plans (VMATauto) were compared with VMAT plans generated manually by 1 expert dosimetrist in the absence of time pressure (VMATman). For all treatment plans, planning target volume (PTV) coverage and sparing of organs-at-risk were quantified.

Results: All generated plans were clinically acceptable and had similar PTV coverage (V95% > 99%). For VMATauto and VMATman plans, the organ-at-risk sparing was similar as well, although only the former plans were generated without any planning workload.

Conclusions: Fully automated generation of high-quality VMAT plans for prostate cancer patients is feasible and has recently been implemented in our clinic.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ijrobp.2013.12.046DOI Listing

Publication Analysis

Top Keywords

fully automated
16
vmat plans
16
prostate cancer
12
cancer patients
12
automated volumetric
8
volumetric modulated
8
modulated arc
8
arc therapy
8
treatment planning
8
automated generation
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!