The accuracy of Vitek MS mass spectrometric identifications was assessed for 206 clinically significant isolates of aerobic Gram-positive bacilli representing 20 genera and 38 species. The Vitek MS identifications were correct for 85% of the isolates (56.3% to the species level, 28.6% limited to the genus level), with misidentifications occurring for 7.3% of the isolates.