During analysis of pure isoprene by gas chromatography/mass spectrometry (GC-MS) using a programmed temperature vaporization (PTV) inlet, the presence of several isoprene dimers was detected in the total ion chromatograms (TICs). This study intends to determine the part of the instrument where dimerization occurs and the relative importance of the dimer amounts under different experimental conditions. The reference thermal dimerization of isoprene gives four six-membered cyclic dimers and two eight-membered ones. In all samples containing different amounts of freshly distilled isoprene, only peaks corresponding to the former appeared in TICs. For the same temperature, their amounts increase as the concentration of injected isoprene increases. The main products are diprene (from 80 to 100%) of the total dimers and dipentene (from 1 to 14%). The sum of the two other dimers is never higher than 6%. In conclusion, isomeric dimers are produced through a dimerization in the inlet. No dimerization of isoprene occurs in the mass spectrometer source. Then care is needed when analyzing terpenic compounds in the presence of isoprene by GC-MS because structures, retention times and mass spectra of diprene and dipentene are close.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.chroma.2014.01.035DOI Listing

Publication Analysis

Top Keywords

dimerization isoprene
12
isoprene
8
isoprene gas
8
gas chromatography/mass
8
chromatography/mass spectrometry
8
presence isoprene
8
dimers
5
unexpected dimerization
4
gas chromatography
4
chromatography inlet
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!