Diagnostic accuracy of computed tomography pulmonary angiography with reduced radiation and contrast material dose: a prospective randomized clinical trial.

Invest Radiol

From the *University Institute of Diagnostic, Interventional and Paediatric Radiology, Inselspital, University Hospital Bern, Berne; †Institute of Radiology, Hospital Centre of Biel, Biel/Bienne, Switzerland; ‡Department of Anaesthesiology and Intensive Care, University of Debrecen, Health and Medical Science Centre, Debrecen, Hungary; §Department of Emergency Medicine, Inselspital, University Hospital Bern, Berne; ∥Department of Emergency Medicine, University Hospital Basel, Basel, Switzerland; ¶Department of Medicine, University of Debrecen, Health and Medical Science Centre, Debrecen, Hungary; and #Department of Radiology, University Hospital Basel, Basel, Switzerland.

Published: April 2014

Objective: The objective of the study was to test the diagnostic performance of low-dose computed tomography pulmonary angiography (CTPA) at peak tube voltage of 80 kVp with both reduced radiation and reduced contrast material (CM) dose.

Materials And Methods: In this single-center, single-blinded prospective randomized trial, 501 patients with body weights of less than 100 kg with suspected acute pulmonary embolism (PE) were assigned to normal-dose CTPA (100-kVp tube energy and 100-mL CM, 255 patients) and low-dose CTPA (80-kVp tube energy and 75-mL CM, 246 patients). Primary end points were evidence of PE in CTPA and accuracy of CTPA on a composite reference standard. Results were compared by calculating the odds ratio with the 95% confidence interval.

Results: The reference diagnosis was equivocal in 20 of the 501 patients. Diagnosis of CTPA was correct in 240 patients and incorrect in 5 in the normal-dose group. Computed tomography pulmonary angiography was correct in 230 patients and incorrect in 6 in the low-dose group (odds ratio, 1.25; 95% confidence interval, 0.38-4.16; P = 0.77). Sensitivity was 96.9% and 100% and specificity was 98.1% and 97.1% in the normal-dose and low-dose groups, respectively. No PE or PE-related death occurred during the 90-day follow-up. The size-specific dose estimates were 30% lower at 80 kVp (4.8 ± 1.0 mGy) compared with that at 100 kVp (6.8 ± 1.2 mGy; P < 0.001).

Conclusions: The accuracy of low-dose CTPA at 80 kVp with a 30% reduced radiation dose and a 25% lower CM volume is not significantly different from that of normal-dose CTPA at 100 kVp in detecting acute PE in patients weighing less than 100 kg.

Download full-text PDF

Source
http://dx.doi.org/10.1097/RLI.0000000000000016DOI Listing

Publication Analysis

Top Keywords

computed tomography
12
tomography pulmonary
12
pulmonary angiography
12
reduced radiation
12
contrast material
8
prospective randomized
8
ctpa
8
501 patients
8
normal-dose ctpa
8
tube energy
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!