Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Purpose: To evaluate the current beliefs about the ways to reduce target registration error (TRE) values in image guided Sinus surgery by rearranging the fiducial configuration, and investigating the best configurations for various surgical fields in a phantom study.
Methods: A new CT-compatible skull phantom consisting of implanted targets was designed to enable direct measurement of TRE in four fields of sinus surgery, Frontal, Ethmoid, Sphenoid and Maxillary. The effects of different landmark configurations on TRE values, measured by the Parsiss-IV navigation system were investigated to find the best landmark arrangement for each region, and compared to the TRE prediction formula to assess the clinically accepted landmark selection approaches based on this formula.
Results: It was shown that smaller values of TRE could be attained by arranging the center of the fiducials to be more focused on the surgery target. The addition of more fiducials and keeping non-linear arrangement of landmark would not necessarily decrease the TRE value.
Conclusion: Optimizing the landmark configuration is important for increasing the localization accuracy in image guided sinus surgery. The common beliefs accepted in the clinical community about the ways to reduce the TRE are very general and should be adapted to specific field of image guided surgery.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jcms.2013.11.019 | DOI Listing |