Background And Purpose: Delta(4) (ScandiDos AB, Uppsala, Sweden) and ArcCHECK with 3DVH software (Sun Nuclear Corp., Melbourne, FL, USA) are commercial quasi-three-dimensional diode dosimetry arrays capable of volumetric measurement-guided dose reconstruction. A method to reconstruct dose for non-coplanar VMAT beams with 3DVH is described. The Delta(4) 3D dose reconstruction on its own phantom for VMAT delivery has not been thoroughly evaluated previously, and we do so by comparison with 3DVH.

Materials And Methods: Reconstructed volumetric doses for VMAT plans delivered with different table angles were compared between the Delta(4) and 3DVH using gamma analysis.

Results: The average γ (2% local dose-error normalization/2mm) passing rate comparing the directly measured Delta(4) diode dose with 3DVH was 98.2 ± 1.6% (1SD). The average passing rate for the full volumetric comparison of the reconstructed doses on a homogeneous cylindrical phantom was 95.6 ± 1.5%. No dependence on the table angle was observed.

Conclusions: Modified 3DVH algorithm is capable of 3D VMAT dose reconstruction on an arbitrary volume for the full range of table angles. Our comparison results between different dosimeters make a compelling case for the use of electronic arrays with high-resolution 3D dose reconstruction as primary means of evaluating spatial dose distributions during IMRT/VMAT verification.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.radonc.2013.12.011DOI Listing

Publication Analysis

Top Keywords

dose reconstruction
20
dose
8
high-resolution dose
8
reconstruction phantom
8
non-coplanar vmat
8
vmat beams
8
table angles
8
passing rate
8
reconstruction
5
vmat
5

Similar Publications

Validity of one-time phantomless patient-specific quality assurance in proton therapy with regard to the reproducibility of beam delivery.

Med Phys

January 2025

OncoRay - National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden, Helmholtz-Zentrum Dresden - Rossendorf, Dresden, Germany.

Background: Patient-specific quality assurance (PSQA) is a crucial yet resource-intensive task in proton therapy, requiring special equipment, expertise and additional beam time. Machine delivery log files contain information about energy, position and monitor units (MU) of all delivered spots, allowing a reconstruction of the applied dose. This raises the prospect of phantomless, log file-based QA (LFQA) as an automated replacement of current phantom-based solutions, provided that such an approach guarantees a comparable level of safety.

View Article and Find Full Text PDF

Rising computed tomography (CT) workloads require more efficient image interpretation methods. Digitally reconstructed radiographs (DRRs), generated from CT data, may enhance workflow efficiency by enabling faster radiological assessments. Various techniques exist for generating DRRs.

View Article and Find Full Text PDF

Alcohol drinking leads to sex-dependent differentiation of T cells.

Eur J Trauma Emerg Surg

January 2025

Department of Trauma Surgery and Orthopedics, Goethe University, University Hospital, Frankfurt, Germany.

Objective: Global per capita alcohol consumption is increasing, posing significant socioeconomic and medical challenges also due to alcohol-related traumatic injuries but also its biological effects. Trauma as a leading cause of death in young adults, is often associated with an increased risk of complications, such as sepsis and multiple organ failure, due to immunological imbalances. Regulatory T cells play a crucial role in maintaining immune homeostasis by regulating the inflammatory response.

View Article and Find Full Text PDF

Background: The removal of preformed antibodies with cleaving enzyme like IdeS (Imlifidase) has demonstrated therapeutic potential in organ transplantation for sensitized recipients. However, preformed xenoreactive antibodies (XAbs) against porcine glycans are predominantly IgM and considered detrimental in pig-to-human xenotransplantation.

Methods: Recombinant IceM, an endopeptidase cleaving IgM, was generated in Escherichia coli.

View Article and Find Full Text PDF

In image-guided radiotherapy (IGRT), four-dimensional cone-beam computed tomography (4D-CBCT) is critical for assessing tumor motion during a patients breathing cycle prior to beam delivery. However, generating 4D-CBCT images with sufficient quality requires significantly more projection images than a standard 3D-CBCT scan, leading to extended scanning times and increased imaging dose to the patient. To address these limitations, there is a strong demand for methods capable of reconstructing high-quality 4D-CBCT images from a 1-minute 3D-CBCT acquisition.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!