Background: Busulfan, an alkylating agent administered prior to hematopoietic stem cell transplantation, has a narrow therapeutic range and wide variability in metabolism. We developed a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for rapid and accurate quantification of plasma busulfan.
Methods: Busulfan was separated and detected using an LC system containing a C18 column equipped with MS/MS. The sample was eluted with a mobile phase gradient for a total run time of 10 min. Plasma busulfan concentration was quantified against a 6-point standard curve in a multiple reaction monitoring mode at mass-to-charge (m/z) 264.1 > 151.1. Precision, recovery, matrix effect, linearity, detection capability, carryover, and stability were evaluated. The range of plasma busulfan concentration was obtained by analyzing samples from 9 children receiving busulfan.
Results: The coefficients of variation of within-run and within-laboratory precision were all below 5%. Recoveries were all within the range of 100-105%. Linearity was verified from 0 to 5,000 ng/mL. Limit of detection and limit of quantification were 1.56 and 25 ng/mL, respectively. Carryover rate was within allowable limits. Plasma busulfan concentration was stable for 2 weeks at -20℃ and -80℃, but decreased by 25% when the plasma was stored for 24 hr at room temperature, and by <5% in 24 hr at 4℃. The plasma busulfan concentrations were between 347 ng/mL and 5,076 ng/mL.
Conclusions: Our method using LC-MS/MS enables highly accurate, reproducible, and rapid busulfan monitoring with minimal sample preparation. The method may also enable safe and proper dosage.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3885779 | PMC |
http://dx.doi.org/10.3343/alm.2014.34.1.7 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!