The N-point moving average (NPMA) is a mathematical low-pass filter that can smooth peaked noninvasively acquired radial pressure waveforms to estimate central aortic systolic pressure using a common denominator of N/4 (where N=the acquisition sampling frequency). The present study investigated whether the NPMA method can be applied to brachial pressure waveforms. In the derivation group, simultaneously recorded invasive high-fidelity brachial and central aortic pressure waveforms from 40 subjects were analyzed to identify the best common denominator. In the validation group, the NPMA method with the obtained common denominator was applied on noninvasive brachial pressure waveforms of 100 subjects. Validity was tested by comparing the noninvasive with the simultaneously recorded invasive central aortic systolic pressure. Noninvasive brachial pressure waveforms were calibrated to the cuff systolic and diastolic blood pressures. In the derivation study, an optimal denominator of N/6 was identified for NPMA to derive central aortic systolic pressure. The mean difference between the invasively/noninvasively estimated (N/6) and invasively measured central aortic systolic pressure was 0.1±3.5 and -0.6±7.6 mm Hg in the derivation and validation study, respectively. It satisfied the Association for the Advancement of Medical Instrumentation standard of 5±8 mm Hg. In conclusion, this method for estimating central aortic systolic pressure using either invasive or noninvasive brachial pressure waves requires a common denominator of N/6. By integrating the NPMA method into the ordinary oscillometric blood pressure determining process, convenient noninvasive central aortic systolic pressure values could be obtained with acceptable accuracy.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1161/HYPERTENSIONAHA.113.02229 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!