Aim: To validate a pretreatment verification method of dose calculation and dose delivery based on measurements with Metaplex PTW phantom.

Background: The dose-response relationships for local tumor control and radiosensitive tissue complications are strong. It is widely accepted that an accuracy of dose delivery of about 3.5% (one standard deviation) is required in modern radiotherapy. This goal is difficult to achieve. This paper describes our experience with the control of dose delivery and calculations at the ICRU reference point.

Materials And Methods: The calculations of dose at the ICRU reference point performed with the treatment planning system CMS XiO were checked by measurements carried out in the PLEXITOM™ phantom. All measurements were performed with the ion chamber positioned in the phantom, at the central axis of the beam, at depth equivalent to the radiological depth (at gantry zero position). The source-to-phantom surface distance was always set to keep the source-to-detector distance equal to the reference point depth defined in the ICRU Report 50 (generally, 100 cm). The dose was measured according to IAEA TRS 398 report for measurements in solid phantoms. The measurement results were corrected with the actual accelerator's output factor and for the non-full scatter conditions. Measurements were made for 111 patients and 327 fields.

Results: The average differences between measurements and calculations were 0.03% (SD = 1.4%), 0.3% (SD = 1.0%), 0.1% (SD = 1.1%), 0.6% (SD = 1.8%), 0.3% (SD = 1.5%) for all measurements, for total dose, for pelvis, thorax and H&N patients, respectively. Only in 15 cases (4.6%), the difference between the measured and the calculated dose was greater than 3%. For these fields, a detailed analysis was undertaken.

Conclusion: The verification method provides an instantaneous verification of dose calculations before the beginning of a patient's treatment. It allows to detect differences smaller than 3.5%.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3863258PMC
http://dx.doi.org/10.1016/j.rpor.2012.12.007DOI Listing

Publication Analysis

Top Keywords

dose delivery
12
dose
10
pretreatment verification
8
verification dose
8
dose calculation
8
measurements
8
plexitom™ phantom
8
verification method
8
icru reference
8
reference point
8

Similar Publications

Objectives: Past studies have shown the efficacy of spinal targeted drug delivery (TDD) in pain relief, reduction in opioid use, and cost-effectiveness in long-term management of complex chronic pain. We conducted a survey to determine treatment variables associated with patient satisfaction.

Materials And Methods: Patients in a single pain clinic who were implanted with Medtronic pain pumps to relieve intractable pain were identified from our electronic health record.

View Article and Find Full Text PDF

Objective: To assess the efficacy of low-dose aspirin in the prevention of adverse outcomes in low-risk, nulliparous singleton pregnancies.

Data Sources: PubMed, Ovid MEDLINE, Scopus, Cochrane Library, clinicaltrials.gov, and ScienceDirect were searched from their inception to August 5, 2023.

View Article and Find Full Text PDF

Breast cancer (BC) is a substantial reason for cancer-related mortality among women across the globe. Anastrozole (ANS) is an effective orally administered hormonal therapy for estrogen+ (ER+) BC treatment. However, several side effects and pharmacokinetic limitations restricted its uses in BC treatment.

View Article and Find Full Text PDF

Purpose: Mesalazine and thiopurines are important therapeutic agents for pediatric patients with ulcerative colitis (UC). Mesalazine, which may be administered in different forms depending on delivery mechanisms, can affect thiopurine metabolism, leading to increased 6-thioguanine nucleotides (6-TGN) levels. Therefore, when using these two drugs simultaneously, their interactions must be considered.

View Article and Find Full Text PDF

Leuprolide acetate is commonly used to reduce the size of myomas before surgery. Initially, it stimulates pituitary gonadotropin secretion, followed by sustained suppression of gonadal function. However, the impact on pregnancy outcomes from inadvertent exposure remains unclear.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!