Validation of an iPad test of letter contrast sensitivity.

Optom Vis Sci

*OD, PhD, FAAO †OD, MS, FAAO ‡OD, FAAO §MCOptom, PhD, FAAO School of Optometry (PSK, MEJ, EJK), Indiana University, Bloomington, Indiana; and College of Optometry (MAB), University of Houston, Houston, Texas.

Published: March 2014

Purpose: An iPad-based letter contrast sensitivity test was developed (ridgevue.com) consisting of two letters on each page of an iBook. The contrast decreases from 80% (logCS = 0.1) to 0.5% (logCS = 2.3) by 0.1 log units per page. The test was compared to the Pelli-Robson Test and the Freiburg Acuity and Contrast Test.

Methods: Twenty normally sighted subjects and 20 low-vision subjects were tested monocularly at 1 m using each test wearing their habitual correction. After a 5-minute break, subjects were retested with each test in reverse order. Two different letter charts were used for both the Pelli-Robson and iPad tests, and the order of testing was varied systematically. For the Freiburg test, the target was a variable contrast Landolt C presented at eight possible orientations and used a 30-trial Best PEST procedure. Repeatability and agreement were assessed by determining the 95% limits of agreement (LoA) ± 1.96 SD of the differences between administrations or tests.

Results: All three tests showed good repeatability in terms of the 95% LoA: iPad = ± 0.19, Pelli-Robson = ± 0.19, and Freiburg = ± 0.15. The iPad test showed good agreement with the Freiburg test with similar mean (± SD) logCS (iPad = 1.98 ± 0.11, Freiburg = 1.96 ± 0.06) and with narrow 95% LoA (± 0.24), but the Pelli-Robson test gave significantly lower values (1.65 ± 0.04). Low-vision subjects had slightly poorer repeatability (iPad = ± 0.24, Pelli-Robson = ± 0.23, Freiburg = ± 0.21). Agreement between the iPad and Freiburg tests was good (iPad = 1.45 ± 0.40, Freiburg = 1.54 ± 0.37), but the Pelli-Robson test gave significantly lower values (1.30 ± 0.30).

Conclusions: The iPad test showed similar repeatability and may be a rapid and convenient alternative to some existing measures. The Pelli-Robson test gave lower values than the other tests.

Download full-text PDF

Source
http://dx.doi.org/10.1097/OPX.0000000000000158DOI Listing

Publication Analysis

Top Keywords

pelli-robson test
16
test
13
ipad test
12
test lower
12
lower values
12
letter contrast
8
contrast sensitivity
8
freiburg
8
low-vision subjects
8
ipad
8

Similar Publications

Purpose: Mutations affecting the CRB1 gene can result in a range of retinal phenotypes, including early onset severe retinal dystrophy/Leber congenital amaurosis (EOSRD/LCA), retinitis pigmentosa, cone-rod dystrophy (CORD), and macular dystrophy (MD). As research into treatment strategies advances towards clinical translation, there is a need to establish reliable outcome metrics. This study explores the contrast sensitivity function (CSF) across different spatial frequencies in individuals with CRB1-retinopathies using the child-friendly PopCSF test, an iPad-based "gamified" assessment.

View Article and Find Full Text PDF

Purpose: Clinical trials of novel therapies for choroideremia require robust and clinically meaningful visual function outcome measures. Best-corrected visual acuity (BCVA) is mostly insensitive to changes in disease state, until late stages, and hence also to potential therapeutic gains after gene therapies. While the insensitivity of BCVA as an effective outcome measure is common wisdom, its low importance has not been rigorously demonstrated in the literature.

View Article and Find Full Text PDF

Lighting modification is commonly performed by optometrists and occupational therapists to enhance visibility and visual comfort among schoolchildren with low vision. The purpose of this study is to determine the optimal illumination level for visual function and visual comfort of schoolchildren with low vision and the relationship between them. A cross-sectional study was conducted to assess five levels of illumination ranging from 125 lux to 2000 lux to determine the optimal illumination for visual functions and visual comfort in schoolchildren with low vision from a special education school for blind in Malaysia.

View Article and Find Full Text PDF

Background/objectives: The use of mobile ophthalmology applications (MOA) is increasing, but many of these tools have not been validated. This study was performed to assess the accuracy of a popular MOA, Eye Handbook, in measuring five commonly-tested vision assessment parameters (distance visual acuity (DVA), near visual acuity (NVA), colour vision testing (CVT), contrast sensitivity (CS), and pupillary distance (PD)) was compared with traditional vision assessment methods (TVAM) [i.e.

View Article and Find Full Text PDF
Article Synopsis
  • SpotChecks is a new self-monitoring contrast sensitivity test, and this study evaluated its reliability compared to in-office tests and Pelli-Robson charts in healthy adults aged 22-84. !* -
  • The study involved 61 participants who underwent multiple tests over a span of days, assessing repeatability using statistical methods and comparing results across different testing environments and age groups. !* -
  • Findings indicated that SpotChecks demonstrated strong repeatability both at home and in the office, with performance slightly declining in individuals aged 50 and above, but overall showing promise as an effective vision self-monitoring tool. !*
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!