Near-infrared spectroscopy (NIRS) is a noninvasive method used to evaluate tissue oxygenation. We evaluated the relationship between cerebral and renal NIRS parameters during transcatheter intervention and adverse events in the catheterization room. Between January 1 and May 31, 2012, 123 of 163 pediatric patients undergoing cardiac catheterization were followed by NIRS. All were monitored by electrocardiography, noninvasive blood pressure measurement, pulse oxymetry, initial and final blood lactate level measurement. The number of interventional procedures was 73 (59%). During the procedures, 39 patients experienced a total of 41 adverse events: 18 (19.5%) had desaturation, 10 (8.1%) arrhythmia, three (2.4%) had respiratory difficulty, six (4.8%) had a situation calling for cardiopulmonary resuscitation, three (2.4%) had anemia necessitating transfusion, and one (0.8%) had a cyanotic spell. Cranial NIRS values worsened in 12 (9.8%) and renal measurements worsened in 13 (12.5%) patients. The sensitivity and specificity of a 9% impairment of cranial values were 90 and 61%, respectively, while the corresponding calculations for a 21% fall in renal measurements were 54% sensitivity and 90% specificity. When arrhythmia developed, NIRS values fell simultaneously, while the development of a desaturation problem was heralded by NIRS falling 10-15 s earlier than changes in pulse oxymetry; on improving saturation, NIRS returned to earlier values 10-15 s before pulse oxymetry readings. NIRS monitoring may provide an early warning with regard to complications likely to develop during a procedure. A fall of 9% in cranial NIRS values, or of 21% in renal measurements, should raise clinician awareness.

Download full-text PDF

Source
http://dx.doi.org/10.1111/aor.12256DOI Listing

Publication Analysis

Top Keywords

pulse oxymetry
12
nirs values
12
renal measurements
12
nirs
9
cardiac catheterization
8
adverse events
8
three 24%
8
cranial nirs
8
values
5
infrared spectroscopy
4

Similar Publications

: Attaining adequate oxygenation in critically ill patients undergoing invasive ventilation necessitates intense monitoring through pulse oximetry (SpO) and frequent manual adjustments of ventilator settings like the fraction of inspired oxygen (FiO) and the level of positive end-expiratory pressure (PEEP). Our aim was to compare the quality of oxygenation with the use of automated ventilation provided by INTELLiVENT-Adaptive Support Ventilation (ASV) vs. ventilation that is not automated, i.

View Article and Find Full Text PDF

Validation of Aerobic Capacity (VO2max) and Pulse Oximetry in Wearable Technology.

Sensors (Basel)

January 2025

School of Integrated Health Sciences, Department of Kinesiology and Nutrition Sciences, University of Nevada, Las Vegas, NV 89154, USA.

Introduction: As wearable technology becomes increasingly popular and sophisticated, independent validation is needed to determine its accuracy and potential applications. Therefore, the purpose of this study was to evaluate the accuracy (validity) of VO2max estimates and blood oxygen saturation measured via pulse oximetry using the Garmin fēnix 6 with a general population participant pool.

Methods: We recruited apparently healthy individuals (both active and sedentary) for VO2max (n = 19) and pulse oximetry testing (n = 22).

View Article and Find Full Text PDF

Accuracy of Rhythm Diagnostic Systems' MultiSense in Detection of Arterial Oxygen Saturation and Respiratory Rate During Hypoxia in Humans: Effects of Skin Color and Device Localization.

Sensors (Basel)

December 2024

Biomedicine Research Center of Strasbourg (CRBS), UR 3072, "Mitochondria, Oxidative Stress and Muscle Plasticity", Faculty of Medicine, University of Strasbourg, 67000 Strasbourg, France.

The continuous monitoring of oxygen saturation (SpO) and respiratory rates (RRs) are major clinical issues in many cardio-respiratory diseases and have been of tremendous importance during the COVID-19 pandemic. The early detection of hypoxemia was crucial since it precedes significant complications, and SpO follow-up allowed early hospital discharge in patients needing oxygen therapy. Nevertheless, fingertip devices showed some practical limitations.

View Article and Find Full Text PDF

Context: Heart rate (HR) is the most vital parameter to assess hemodynamic transition at birth. ECG is considered a gold standard for HR assessment. New devices with dry electrodes are easy to apply on a wet newborn.

View Article and Find Full Text PDF

Introduction: Hereditary hemorrhagic telangiectasia (HHT) is an autosomal dominant vascular disease and screening to detect pulmonary arteriovenous malformations (PAVMs) is important to prevent complications. In adults, transthoracic contrast echocardiogram (TTCE) is used to screen PAVMs. In children, a conservative screening method seems to be sufficient to rule out major PAVMs and prevent them from PAVM-related complications.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!