Collisions between cold cesium atoms and background gas atoms at ambient temperature reduce the cold atom signal in a fountain clock and at the same time produce a shift in the measured clock frequency. We evaluate the shift in the NPL-CsF2 cesium fountain primary frequency standard based on measurements of the fractional loss of cold atoms from the atomic cloud during the interrogation time combined with a model by Gibble that quantifies the relationship between the loss and the frequency shift.

Download full-text PDF

Source
http://dx.doi.org/10.1109/TUFFC.2014.6689789DOI Listing

Publication Analysis

Top Keywords

frequency shift
8
background gas
8
primary frequency
8
frequency standard
8
evaluation frequency
4
shift
4
shift caused
4
caused collisions
4
collisions background
4
gas primary
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?