Impaction grafted bone chip size effect on initial stability in an acetabular model: Mechanical evaluation.

J Orthop

Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds LS2 9JT, UK ; School of Mechanical Engineering, Xian Jiaotong University, Shanghai 200240, China.

Published: January 2014

Introduction: Acetabular bone defect reconstruction is an increasing problem for surgeons with patients undergoing complex primary or revision total hip replacement surgery. Impaction bone grafting is one technique that has favourable long-term clinical outcome results for patients who undergo this reconstruction method for acetabular bone defects. Creating initial mechanical stability of the impaction bone graft in this technique is known to be the key factor in achieving a favourable implant survival rate. Different sizes of bone chips were used in this technique to investigate if the size of bone chips used affected initial mechanical stability of a reconstructed acetabulum.

Methodology: Twenty acetabular models were created in total. Five control models were created with a cemented cup in a normal acetabulum. Then five models in three different groups of bone chip size were constructed. The three groups had an acetabular protrusion defect reconstructed using either; 2-4 mm(3), 10 mm(3) or 20 mm(3) bone chip size for impaction grafting reconstruction. The models underwent compression loading up to 9500 N and displacement within the acetabular model was measured indicating the initial mechanical stability.

Results: This study reveals that, although not statistically significant, the largest (20 mm(3)) bone chip size grafted models have an inferior maximum stiffness compared to the medium (10 mm(3)) bone chip size.

Interpretations: Our study suggests that 10 mm(3) size of bone chips provide better initial mechanical stability compared to smaller or larger bone chips. We dismissed the previously held opinion that the biggest practically possible graft is best for acetabular bone graft impaction.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3849241PMC
http://dx.doi.org/10.1016/j.jor.2013.09.004DOI Listing

Publication Analysis

Top Keywords

bone chip
20
chip size
16
initial mechanical
16
bone chips
16
bone
14
acetabular bone
12
mechanical stability
12
acetabular model
8
impaction bone
8
bone graft
8