We have previously reported an 18-year-old girl with a congenital heart defect who developed complete heart block after one of her corrective surgeries and who needed an epicardial pacemaker implantation. She developed contact sensitivity to silicone compounds. The problem was solved by implanting a silicone-free pacemaker system utilizing silicone-free transvenous leads. The patient was readmitted 2 years later due to lead failure. As no silicone-free epicardial leads were available, we decided to use standard silicone epicardial leads and enclose the whole system in Gore-Tex material (W.L. Gore & Associates, Flagstaff, AZ). Based on our experience we would discourage the use of silicone-free transvenous pacing leads for epicardial use.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.athoracsur.2013.04.135 | DOI Listing |
Ann Thorac Surg
January 2014
Department of Pediatric Cardiac Surgery, Rheinisch-Westfaelische Technische Hochschule Aachen University, Aachen, Germany.
We have previously reported an 18-year-old girl with a congenital heart defect who developed complete heart block after one of her corrective surgeries and who needed an epicardial pacemaker implantation. She developed contact sensitivity to silicone compounds. The problem was solved by implanting a silicone-free pacemaker system utilizing silicone-free transvenous leads.
View Article and Find Full Text PDFAnn Thorac Surg
April 2009
Department of Pediatric Cardiac Surgery, RWTH Aachen University, Aachen, Germany.
Silicone is a widely used biomaterial. Contact allergy, particularly to silicone components of pacemaker coatings, is uncommon. We present a 12-year-old girl with a history of complex congenital heart disease and acquired complete heart block excluding transvenous lead placement.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!