Validation of an ultrasound dilution technology for cardiac output measurement and shunt detection in infants and children.

Pediatr Crit Care Med

1Department of Anesthesia and Intensive Care, Children's Hospital, Pediatric Intensive Care Unit, University Hospital of Lund, Lund, Sweden. 2Department of Pediatric Cardiac Surgery, Children's Hospital, University Hospital of Lund, Lund, Sweden.

Published: February 2014

Objective: To validate cardiac output measurements by ultrasound dilution technology (COstatus monitor) against those obtained by a transit-time ultrasound technology with a perivascular flow probe and to investigate ultrasound dilution ability to estimate pulmonary to systemic blood flow ratio in children.

Design: Prospective observational clinical trial.

Setting: Pediatric cardiac operating theater in a university hospital.

Material And Methods: In 21 children (6.1 ± 2.6 kg, mean ± SD) undergoing heart surgery, cardiac output was simultaneously recorded by ultrasound dilution (extracorporeal arteriovenous loop connected to existing arterial and central venous catheters) and a transit-time ultrasound probe applied to the ascending aorta, and when possible, the main pulmonary artery. The pulmonary to systemic blood flow ratio estimated from ultrasound dilution curve analysis was compared with that estimated from transit-time ultrasound technology.

Results: Bland-Altman analysis of the whole cohort (90 pairs, before and after surgery) showed a bias between transit-time ultrasound (1.01 ± 0.47 L/min) and ultrasound dilution technology (1.03 ± 0.51 L/min) of -0.02 L/min, limits of agreement -0.3 to 0.3 L/min, and percentage error of 31%. In children with no residual shunts, the bias was -0.04 L/min, limits of agreement -0.28 to 0.2 L/min, and percentage error 19%. The pooled co efficient of variation was for the whole cohort 3.5% (transit-time ultrasound) and 6.3% (ultrasound dilution), and in children without shunt, it was 2.9% (transit-time ultrasound) and 4% (ultrasound dilution), respectively. Ultrasound dilution identified the presence of shunts (pulmonary to systemic blood flow ≠ 1) with a sensitivity of 100% and a specificity of 92%. Mean pulmonary to systemic blood flow ratio by transit-time ultrasound was 2.6 ± 1.0 and by ultrasound dilution 2.2 ± 0.7 (not significant).

Conclusion: The COstatus monitor is a reliable technique to measure cardiac output in children with high sensitivity and specificity for detecting the presence of shunts.

Download full-text PDF

Source
http://dx.doi.org/10.1097/PCC.0000000000000053DOI Listing

Publication Analysis

Top Keywords

ultrasound dilution
40
transit-time ultrasound
28
cardiac output
16
ultrasound
16
pulmonary systemic
16
systemic blood
16
blood flow
16
dilution technology
12
flow ratio
12
dilution
10

Similar Publications

In recent years, significant advancements in respiratory and nutritional care have markedly improved the survival rates of preterm infants and enhanced long-term health outcomes. Despite these improvements, emerging research highlights the lasting impacts of early growth patterns on an individual's health trajectory. Adults born prematurely face a higher incidence of health issues related to their early birth.

View Article and Find Full Text PDF

Aims: Coronary microvascular dysfunction (CMD) is a heterogeneous condition defined by reduced coronary flow reserve (CFR). The new index 'microvascular resistance reserve' (MRR) has been developed, but its role is unclear. We investigate the relationships between functional indices in ANOCA (angina and non-obstructive coronary arteries) patients and evaluate the hemodynamic features of different CMD subtypes.

View Article and Find Full Text PDF

The cooling procedure markedly diminishes the quality of guinea pig () sperms, primarily because their membranes are highly susceptible to this process. This susceptibility triggers the generation of reactive oxygen species and free radicals, ultimately leading to lipid peroxidation in the sperm membrane. Surprisingly, there has been a lack of research on the use of Tris-based extenders to safeguard guinea pig sperm under refrigeration conditions.

View Article and Find Full Text PDF

Objective: To assess the agreement of transpulmonary thermodilution (TPTD) and transpulmonary ultrasound dilution (TPUD) against direct measurement of main pulmonary artery flow with an ultrasound transit time flow probe (UTF) over a wide range of conditions in anesthetized cats. Additionally, the trending ability of TPTD, TPUD, and esophageal Doppler ultrasonography (EDU) was evaluated against UTF.

Methods: 12 purpose-bred cats were used.

View Article and Find Full Text PDF

Background: Colorectal obstruction is a critical condition requiring prompt diagnosis and intervention. Gastrografin, a water-soluble contrast agent, combines diagnostic and therapeutic benefits, facilitating bowel cleansing and enhancing intestinal motility. This study assessed the safety and effectiveness of Gastrografin enemas in emergency settings.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!