Performance of an automated multiplex immunofluorescence assay for detection of Chlamydia trachomatis immunoglobulin G.

Diagn Microbiol Infect Dis

Center for Research on Intracellular Bacteria, Institute of Microbiology, Faculty of Biology and Medicine, University of Lausanne and University Hospital Center (CHUV), Lausanne, Switzerland. Electronic address:

Published: March 2014

Chlamydia serology is indicated to investigate etiology of miscarriage, infertility, pelvic inflammatory disease, and ectopic pregnancy. Here, we assessed the reliability of a new automated-multiplex immunofluorescence assay (InoDiag test) to detect specific anti-C. trachomatis immunoglobulin G. Considering immunofluorescence assay (IF) as gold standard, InoDiag tests exhibited similar sensitivities (65.5%) but better specificities (95.1%-98%) than enzyme-linked immunosorbent assays (ELISAs). InoDiag tests demonstrated similar or lower cross-reactivity rates when compared to ELISA or IF.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.diagmicrobio.2013.11.022DOI Listing