Purpose: The aim of this work was to extend an in-vivo dosimetry (IVD) method, previously developed by the authors for 3D-conformal radiotherapy, to step and shoot IMRT treatments for pelvic tumors delivered by Elekta linacs.
Materials And Methods: The algorithm is based on correlation functions to convert EPID transit signals into in-vivo dose values at the isocenter point, Diso. The EPID images were obtained by the so-called "IMRT Dosimetric Weighting" mode as a superposition of many segment fields. This way each integral dosimetric image could be acquired in about 10 s after the end of beam delivery and could be processed while delivering the successive IMRT beams. A specific algorithm for Diso reconstruction especially featured for step and shoot IMRT was implemented using a fluence inhomogeneity index, FI, introduced to describe the degree of beam modulation with respect to open beams. A γ-analysis of 2D-EPID images obtained day to day, resulted rapid enough to verify the plan delivery reproducibility.
Results: Fifty clinical IMRT beams, planned for patients undergoing radiotherapy of pelvic tumors, were used to irradiate a homogeneous phantom. For each beam the agreement between the reconstructed dose, Diso, and the TPS computed dose, Diso,TPS, was well within 5%, while the mean ratio R = Diso/Diso,TPS resulted for 250 tests equal to 1.006 ± 0.036. The same beams were checked in vivo, i.e. during patient treatment delivery, obtaining 500 tests whose average R ratio resulted equal to 1.011 ± 0.042. The γ-analysis of the EPID images with 5% 3 mm criteria supplied 85% of the tests with pass rates γ(mean) ≤ 0.5 and P(γ<1) ≥ 90%.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ejmp.2013.11.005 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!