Comparison of the effect of iterative reconstruction versus filtered back projection on cardiac CT postprocessing.

Acad Radiol

Heart and Vascular Center, Medical University of South Carolina, Ashley River Tower, 25 Courtenay Drive, Charleston, SC, 29425-2260; Department of Cardiology and Intensive Care Medicine, Heart Center Munich-Bogenhausen, Munich, Germany.

Published: March 2014

Rationale And Objectives: To investigate the impact of iterative reconstruction in image space (IRIS) on image noise, image quality (IQ), and postprocessing at coronary computed tomography angiography (cCTA) compared to traditional filtered back-projection (FBP).

Materials And Methods: The cCTA results of 50 patients (26 men; 58 ± 15 years, body mass index 31.5 ± 6.7 kg/m²) were investigated using a second-generation dual-source computed tomography system. Scan data were reconstructed with the use of IRIS and FBP algorithms. Two radiologists independently evaluated the reconstructions using automated coronary tree analysis software. Image noise was measured and IQ was rated on a 5-point Likert scale. The number of manual corrections after automated vessel segmentation, the time required to complete segmentation, and the number of missed segments were assessed in both IRIS and FBP reconstructions. Results were compared using paired t-test.

Results: IRIS significantly reduced image noise compared to FBP (23.3 ± 8.8 vs. 33.5 ± 13.5 Hounsfield units, P < .001). Subjective IQ improved with IRIS (IRIS 3.2 ± 1.0 vs. FBP 3.0 ± 1.0, P < .05). IRIS decreased the time needed for coronary segmentation from 111.9 ± 40.5 seconds to 95.2 ± 38.2 seconds with FBP (P < .01) and required fewer manual corrections (5.7 ± 3.0 vs. 6.8 ± 3.6, P < .01). The number of missed vessel segments was not significantly different (3.6 ± 1.8 vs. 3.8 ± 1.9, P > .05) between IRIS and FBP, respectively.

Conclusions: During cCTA postprocessing, IRIS significantly decreases the time and the number of manual corrections for a complete coronary segmentation compared to FBP. This effect is likely attributable to suppression of image noise by IRIS, which improves the performance of automated vessel segmentation and positively impacts cCTA analysis.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.acra.2013.11.008DOI Listing

Publication Analysis

Top Keywords

image noise
16
iris fbp
16
manual corrections
12
iris
10
iterative reconstruction
8
computed tomography
8
number manual
8
automated vessel
8
vessel segmentation
8
number missed
8