Light scattering and light transmittance of cadaver eye-explanted intraocular lenses of different materials.

J Cataract Refract Surg

From the Department of Ophthalmology and Visual Sciences (Morris, Werner, Liu, Stallings, Floyd), John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA; and Visão Institutos Oftalmológicos Associados (Barra), Brasília, DF, Brazil.

Published: January 2014

Purpose: To evaluate light scattering and light transmittance in cadaver eye-explanted intraocular lenses (IOLs) manufactured from different materials.

Setting: John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA.

Design: Experimental study.

Methods: Forty-nine pseudophakic cadaver eyes were selected according to IOL material/type and implantation duration, and the IOLs were explanted. Hydrophobic acrylic, hydrophilic acrylic, poly(methyl methacrylate) (PMMA), and silicone IOLs were included. Gross and light microscopy was performed for all IOLs. Light scattering was measured with an EAS 1000 Scheimpflug camera, and light transmittance was assessed using a Lambda 35 UV/Vis spectrophotometer (single-beam configuration with an RSA PE-20 integrating sphere). Analyses were performed at room temperature in the hydrated state and compared with analyses of controls.

Results: The highest levels of surface light scattering were measured for 3-piece hydrophobic acrylic, which was also the IOL type with the longest implantation duration among the Acrysof hydrophobic acrylic IOLs. Hydrophilic acrylic, PMMA, and silicone IOLs exhibited relatively low light-scattering levels. The lowest light-scattering levels were observed with PMMA IOLs (1-piece looped and 3-piece) and plate silicone IOLs, which represent the IOL types with the longest implantation duration in this series. Light transmittance values measured for all IOL types appeared to be similar to the values of the corresponding control IOLs.

Conclusions: The phenomenon of surface light scattering (nanoglistenings) is more particularly related to hydrophobic acrylic IOLs and increases with implantation time. No significant effect of surface light scattering on IOL light transmittance was found.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcrs.2013.10.016DOI Listing

Publication Analysis

Top Keywords

light scattering
24
light transmittance
20
hydrophobic acrylic
16
light
12
implantation duration
12
silicone iols
12
surface light
12
iols
9
scattering light
8
transmittance cadaver
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!