A high dynamic range pulse counting detection system for mass spectrometry.

Rapid Commun Mass Spectrom

AB Sciex, 71 Four Valley Dr., Concord, Ontario, L4K 4V8, Canada.

Published: January 2014

Rationale: A high dynamic range pulse counting system has been developed that demonstrates an ability to operate at up to 2e8 counts per second (cps) on a triple quadrupole mass spectrometer. Previous pulse counting detection systems have typically been limited to about 1e7 cps at the upper end of the systems dynamic range. Modifications to the detection electronics and dead time correction algorithm are described in this paper.

Methods: A high gain transimpedance amplifier is employed that allows a multi-channel electron multiplier to be operated at a significantly lower bias potential than in previous pulse counting systems. The system utilises a high-energy conversion dynode, a multi-channel electron multiplier, a high gain transimpedance amplifier, non-paralysing detection electronics and a modified dead time correction algorithm. Modification of the dead time correction algorithm is necessary due to a characteristic of the pulse counting electronics.

Results: A pulse counting detection system with the capability to count at ion arrival rates of up to 2e8 cps is described. This is shown to provide a linear dynamic range of nearly five orders of magnitude for a sample of aprazolam with concentrations ranging from 0.0006970 ng/mL to 3333 ng/mL while monitoring the m/z 309.1 → m/z 205.2 transition. This represents an upward extension of the detector's linear dynamic range of about two orders of magnitude.

Conclusions: A new high dynamic range pulse counting system has been developed demonstrating the ability to operate at up to 2e8 cps on a triple quadrupole mass spectrometer. This provides an upward extension of the detector's linear dynamic range by about two orders of magnitude over previous pulse counting systems.

Download full-text PDF

Source
http://dx.doi.org/10.1002/rcm.6777DOI Listing

Publication Analysis

Top Keywords

pulse counting
32
dynamic range
28
high dynamic
12
range pulse
12
counting detection
12
previous pulse
12
dead time
12
time correction
12
correction algorithm
12
linear dynamic
12