Purpose: To determine the feasibility of Ultra wide-field intravenous fundus fluorescein angiography (UWF-IV-FFA) in infants.

Methods: A retrospective review of infants who underwent UWF-IV-FFA with the Optos P200MA was performed. The ability to capture different phases of the angiogram, the total image acquisition time and the presence of image artefacts were analysed.

Results: Twelve infants with a variety of proliferative retinopathies underwent UWF-IV-FFA over a 7-month period. The mean age was 3.4 months (range 2-6 months) and the mean image acquisition time was 4 min (range 3-5). Pseudocolour fundus images and the venous, recirculation and late phases of UWF-IV-FFA were captured successfully in all infants (100%). Choroidal and arterial phases were captured in one (8.3%) and 10 (83.3%) infants, respectively. Image artefacts due to eyelashes and corneal desiccation occurred in nine (75%) and six (50%) infants, respectively. No adverse medical problems were noted.

Conclusions: We have shown that UWF-IV-FFA can be successfully and safely performed in infants using the Optos P200MA system. It is a feasible new alternative to RetCam for obtaining an IV-FFA in the outpatients setting.

Download full-text PDF

Source
http://dx.doi.org/10.1136/bjophthalmol-2013-304450DOI Listing

Publication Analysis

Top Keywords

optos p200ma
12
ultra wide-field
8
wide-field intravenous
8
intravenous fundus
8
fundus fluorescein
8
fluorescein angiography
8
infants optos
8
underwent uwf-iv-ffa
8
image acquisition
8
acquisition time
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!