Background: In patients with normal left ventricular (LV) ejection fraction (EF), the interposition of chronic, high-dose right ventricular apical (RVA) pacing may produce late EF decline.

Objective: To test the hypothesis that LV dyssynchrony, defined echocardiographically and apparent early after interposition of pacing, would be greater in patients who subsequently demonstrated EF decline.

Methods: Ninety-one patients with normal prepacing EF who underwent atrioventricular node ablation and subsequent high-dose RVA pacing were studied. Transthoracic echocardiograms were performed early (median 4 months) and late (median 28 months) after interposition of pacing, with a significant decline in EF between these studies defined as ≥5%. Speckle-tracking longitudinal strain analysis of the early echocardiogram was performed to quantify dyssynchrony. In addition to standard dyssynchrony indices, a novel index of apex-to-base mechanical propagation delay (MPD) was used.

Results: Multivariable analysis determined that MPD of the septum correlated with a significant decline in EF, independent of all other dyssynchrony, clinical, or pacing variables. A septal MPD value exceeding 50 ms was associated with EF decline at 81% sensitivity and 88% specificity.

Conclusions: Dyssynchrony, in particular septal MPD, measured early after interposition of high-dose RVA pacing predicted a significant late decline in EF in patients who had normal prepacing EF.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hrthm.2013.12.020DOI Listing

Publication Analysis

Top Keywords

patients normal
12
rva pacing
12
ventricular apical
8
left ventricular
8
ejection fraction
8
early interposition
8
interposition pacing
8
normal prepacing
8
high-dose rva
8
median months
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!