Temperature-density hysteresis in X-ray CT during HIFU thermal ablation: heating and cooling phantom study.

Int J Hyperthermia

Department of Biomedical Engineering, Technion IIT , Haifa , Israel .

Published: February 2014

AI Article Synopsis

  • The study explored how thermal ablation affects Hounsfield units (HU) observed in X-ray CT scans, aiming to monitor tissue changes during heating.
  • Eight phantoms and 15 ex vivo tissue samples were subjected to high intensity focused ultrasound (HIFU) at temperatures exceeding 50 °C, with simultaneous CT scanning and temperature monitoring.
  • Results showed an exponential increase in HU during heating and linear behavior during cooling, indicating a hysteresis effect that correlates with maximum temperature, suggesting that this measurable change could serve as a non-invasive method for assessing tissue damage.

Article Abstract

Purpose: This paper investigated the effects of thermal ablation treatment on imaged X-ray computed tomography (CT) Hounsfield units (HU), for the purpose of monitoring tissue denaturation and coagulation.

Materials And Methods: Eight phantoms of water, oil, and chicken serum albumin as well as 15 ex vivo tissue samples were heated by applying high intensity focused ultrasound (HIFU) for 10 to 29 min to obtain denaturation temperatures, (i.e. >50 °C). X-ray CT scanning was performed simultaneously during heating and post-ablation cooling stages, and the HU at the focal zone were registered. The temperature profile versus time was also monitored under similar conditions using a thermocouple probe. The results were plotted and correlated as curves of HU versus temperature.

Results: In all specimens studied, HU values depicted an exponential curve as a function of temperature during the heating stage. However, linear behaviour was observed during the cool-down stage for both chicken serum albumin and ex vivo bovine liver. Thus, a hysteresis phenomenon occurred only when the thermal conditions induced irreversible changes in the sample with quantification demonstrating high correlation with the maximal temperature reached during treatment (R(2)> 0.9) for the chicken serum albumin.

Conclusions: Our results demonstrate a HU-temperature hysteresis phenomenon for HIFU ablation, which is detectible by X-ray CT. This hysteresis is related to the amount of heat induced into the tissue and could potentially indicate irreversible tissue damage. Accordingly, this measurable phenomenon can be utilised as a quantitative method for non-invasive monitoring of thermal ablation.

Download full-text PDF

Source
http://dx.doi.org/10.3109/02656736.2013.860241DOI Listing

Publication Analysis

Top Keywords

thermal ablation
12
chicken serum
12
serum albumin
8
hysteresis phenomenon
8
temperature-density hysteresis
4
x-ray
4
hysteresis x-ray
4
x-ray hifu
4
thermal
4
hifu thermal
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!