Cochlear implants (CI) have now become a standard method of treating severe to profound hearing loss. Recently, the number of patients with CI has been rapidly increasing as the big benefits of CI become more widely known. Magnetic resonance imaging (MRI) has also become a routine diagnostic imaging modality, used in the diagnosis of common conditions, including stroke, back pain, and headache. We report our recent experience with a case in which internal magnet of the cochlear implant was reversed after 1.5-T lumbar spine MRI. This complication is managed successfully by reversing the orientation of the external magnet in the head coil.