Introduction: Our objective was to evaluate the feasibility and initial performance of Backstop reverse thermosensitive polymer to prevent antegrade stone fragment migration during percutaneous nephrolithotomy.

Technical Considerations: Backstop was used during 5 percutaneous nephrolithotomy procedures to prevent stone migration into the ureter. Backstop was placed into the upper ureter under direct vision with a flexible nephroscope or with fluoroscopic guidance. Ultrasonic lithotripsy was then performed. Each patient was evaluated for the following: stone-free rate (postoperative computed tomography), rate of antegrade fragment migration, need for subsequent procedures, and complication rate. The average stone burden was 806 mm(2). Backstop was successfully deployed in all cases. Average procedure length was 106 minutes. Three patients were rendered stone free, 1 patient retained 2 1-mm fragments in the kidney, and 1 patient retained multiple fragments in the kidney, distal ureter, and bladder, all of which were <3 mm. No patient required a secondary procedure, and there were no adverse events.

Conclusion: Backstop appears to be feasible, safe, and easy to use. Backstop should be administered every 45 minutes for longer cases, and warm saline should be used during the case to prevent dissolution. A randomized controlled trial with a larger study population is needed to further evaluate these initial findings.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.urology.2013.10.005DOI Listing

Publication Analysis

Top Keywords

fragment migration
12
percutaneous nephrolithotomy
8
patient retained
8
fragments kidney
8
antegrade reverse-thermosensitive
4
reverse-thermosensitive polymer
4
polymer percutaneous
4
nephrolithotomy prevent
4
prevent fragment
4
migration
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!