Background: The SphygmoCor is used for noninvasive assessment of ascending aortic blood pressure (BP). However, the validity of the SphygmoCor transfer function has not been tested in an exclusively type 2 diabetic patient sample. Calibration with systolic (SBP) and diastolic (DBP) brachial BP has previously been associated with substantial imprecision of central BP estimates. We hypothesized that different noninvasive calibration strategies might improve the accuracy of the estimated ascending aortic BPs.

Methods: In 34 patients with type 2 diabetes we estimated ascending aortic SBP and DBP using the SphygmoCor device and compared these data with invasively recorded data. The validity of the transfer function was assessed by calibrating with invasively recorded DBP and mean BP (MBP). The influence of noninvasive calibration strategies was assessed by calibrating with brachial oscillometric SBP+DBP vs. DBP+MBP using a form factor (ff) of 0.33 and 0.40, respectively.

Results: When calibrating with invasive BP, the difference between estimated and invasively measured ascending aortic SBP and DBP was -2.3±5.6/1.0±0.9 mm Hg. When calibrating with oscillometric brachial BPs, the differences were -9.6±8.1/14.1±6.2 mm Hg (calibration with SBP and DBP), -8.3±11.7/13.9±6.1 mm Hg (DBP and MBP; ff = 0.33), and 1.9±12.2/14.1±6.2 mm Hg (DBP and MBP; ff = 0.40), respectively. Calibration with the average of 3 brachial BPs did not improve accuracy.

Conclusions: The SphygmoCor transfer function seems valid in patients with type 2 diabetes. Noninvasive calibration with DBP and MBP (ff = 0.40) enables accurate estimation of mean ascending aortic SBP at the group level. However, the wide limits of agreement indicate limited accuracy in the individual patient.

Clinical Trials Registration: Clinical Trials No. NCT01538290.

Download full-text PDF

Source
http://dx.doi.org/10.1093/ajh/hpt195DOI Listing

Publication Analysis

Top Keywords

ascending aortic
20
dbp mbp
16
patients type
12
type diabetes
12
transfer function
12
noninvasive calibration
12
aortic sbp
12
sbp dbp
12
blood pressure
8
invasively measured
8

Similar Publications

Aims: This study aims to verify the feasibility and safety of percutaneous coronary intervention (PCI) after a distal transradial approach (dTRA) with radial artery occlusion (RAO) recanalization.

Methods: Between July 2018 and January 2022, 30 patients underwent PCI following attempted RAO recanalization via dTRA. Among these cases, the target radial arteries could not be recanalized in five patients, necessitating alternative vascular access.

View Article and Find Full Text PDF

Ortner's syndrome, also known as cardiovocal syndrome, is a rare cause of hoarseness due to compression of the left recurrent laryngeal nerve caused by pathology of cardiovascular structures in the mediastinum. It was first described by Norbert Ortner in 1897, who associated the syndrome with mitral stenosis. It typically presents as paresis of the left recurrent laryngeal nerve, which is mechanically compressed in the area of the aortic arch.

View Article and Find Full Text PDF

The Nonsyndromic Ascending Thoracic Aorta in a Population-Based Setting: A 5-Year Prospective Cohort Study.

J Am Coll Cardiol

November 2024

Elite Centre for Individualized Medicine in Arterial Disease, Odense University Hospital, Odense, Denmark; Department of Cardiothoracic and Vascular Surgery, Odense University Hospital, Odense, Denmark; Department of Clinical Research, University of Southern Denmark, Odense, Denmark.

Background: Prospective data on the clinical course of the ascending thoracic aorta are lacking.

Objectives: This study sought to estimate growth rates of the ascending aorta and to evaluate occurrences of adverse aortic events (AAEs)-that is, thoracic aortic ruptures, type A aortic dissections, and thoracic aortic-related deaths.

Methods: In this prospective cohort study from the population-based, multicenter, randomized DANCAVAS (Danish Cardiovascular Screening trials) I and II, participants underwent cardiovascular risk assessments including electrocardiogram-gated, noncontrast computed tomography (CT) scans.

View Article and Find Full Text PDF

Does Ascending Aortic Size Matter?: A Plea for Patient-Specific Surveillance and Risk Estimation.

J Am Coll Cardiol

December 2024

Royal Prince Alfred Hospital, Sydney, New South Wales, Australia; Sydney Adventist Hospital, Sydney, New South Wales, Australia; Maquarie University Hospital, Sydney, New South Wales, Australia; Sydney Medical School, University of Sydney, Sydney, New South Wales, Australia.

View Article and Find Full Text PDF

The three-vessel view (3VV) is a standardized transverse scanning plane used in fetal cardiac ultrasound screening to measure the absolute and relative diameters of the pulmonary artery (PA), ascending aorta (Ao), and superior vena cava, as required. The PA/Ao ratio is used to support the diagnosis of congenital heart disease (CHD). However, vascular diameters are measured manually by examiners, which causes intra- and interobserver variability in clinical practice.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!