Background: We set out to develop normal databases and prospectively validate abnormality criteria for a low-dose Tc-99m tetrofosmin myocardial perfusion SPECT protocol using the 530c CZT camera.

Methods: All patients received 6 mCi rest/20 mCi stress doses of Tc-99m tetrofosmin. Rest and stress images were obtained over 7-9 and 5-7 minutes according to the chest size. Low-dose CT of the chest was obtained on a standalone CT scanner. Forty patients with very low likelihood (LLK) of coronary artery disease (CAD) were used to define the normal count distributions. The abnormality criteria were prospectively validated in 55 patients who had coronary angiography and in 40 patients with LLK of CAD.

Results: The results for quantitative non-attenuation-corrected (AC) and AC analysis and visual analysis were as follows: sensitivity of 79%, 85%, and 92% (P = NS) and specificity of 44%, 75%, and 56% (P = NS), respectively. The normalcy rates for quantitative non-AC and AC analyses and visual analysis were 95%, 98%, and 98% (P = NS).

Conclusions: We have developed non-AC and AC normal databases for low-dose rest/stress Tc-99m tetrofosmin myocardial perfusion SPECT protocol using the 530c CZT camera. The per-patient diagnostic performance of quantitative analyses is not significantly different from visual analysis by an experienced reader.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s12350-013-9827-7DOI Listing

Publication Analysis

Top Keywords

tc-99m tetrofosmin
16
visual analysis
16
tetrofosmin myocardial
12
myocardial perfusion
12
perfusion spect
12
530c czt
12
diagnostic performance
8
low-dose rest/stress
8
rest/stress tc-99m
8
czt camera
8

Similar Publications

Objective: Patients with chronic kidney disease (CKD) have an increased risk of adverse cardio-cerebrovascular events. The purpose of this study is to evaluate the prognostic predictors over 5 years in patients with CKD including haemodialysis.

Methods: In this multicenter, prospective cohort study performed with the Gunma-CKD SPECT Study protocol, 311 patients with CKD [estimated glomerular filtration rate (eGFR) < 60 min/ml/1.

View Article and Find Full Text PDF

Various techniques have been used in attempts to reduce interfering gastrointestinal activity in myocardial perfusion imaging (MPI); however, these approaches have yielded inconsistent results. The goal of this study was to investigate the efficacy of monitored walking, a previously unexplored technique, in reducing subdiaphragmatic activity-related artifacts during pharmacologic stress Tc-tetrofosmin MPI with SPECT to improve the overall image quality. The study included patients who underwent MPI with pharmacologic stress.

View Article and Find Full Text PDF

Myocardial perfusion imaging (MPI) provides physiologic and functional information about the heart muscle and its blood flow. Extracardiac radioactivity can interfere with visualization of the inferior wall of the myocardium, leading to poor-quality images, difficulties in interpretation, and delays in routine practice. This study aimed to identify the efficiency of having the patient consume a carbonated lemon drink to minimize the extracardiac radioactivity of Tc-sestamibi in comparison to Tc-tetrofosmin during MPI.

View Article and Find Full Text PDF