Biomechanical analysis of an interspinous fusion device as a stand-alone and as supplemental fixation to posterior expandable interbody cages in the lumbar spine.

J Neurosurg Spine

H. Lee Moffitt Cancer Center & Research Institute, Neuro-Oncology Program and Department of Neurosurgery and Orthopedics, Morsani College of Medicine, University of South Florida;

Published: February 2014

Object: In this paper the authors evaluate through in vitro biomechanical testing the performance of an interspinous fusion device as a stand-alone device, after lumbar decompression surgery, and as supplemental fixation to expandable cages in a posterior lumbar interbody fusion (PLIF) construct.

Methods: Nine L3-4 human cadaveric spines were biomechanically tested under the following conditions: 1) intact/control; 2) L3-4 left hemilaminotomy with partial discectomy (injury); 3) interspinous spacer (ISS); 4) bilateral pedicle screw system (BPSS); 5) bilateral hemilaminectomy, discectomy, and expandable posterior interbody cages with ISS (PLIF-ISS); and 6) PLIF-BPSS. Each test consisted of 100 N of axial preload with ± 7.5 Nm of torque in flexion-extension, right/left lateral bending, and right/left axial rotation. Significant changes in range of motion (ROM), neutral zone stiffness (NZS), elastic zone stiffness (EZS), and energy loss (EL) were explored among conditions using nonparametric Friedman test and Wilcoxon signed-rank comparisons (p ≤ 0.05).

Results: The injury increased ROM in flexion (p = 0.01), left bending (p = 0.03), and right/left rotation (p < 0.01) and also decreased NZS in flexion (p = 0.01) and extension (p < 0.01). Both the ISS and BPSS reduced flexion-extension ROM and increased flexion-extension stiffness (NZS and EZS) with respect to the injury and intact conditions (p < 0.05), but the ISS condition provided greater resistance than BPSS in extension for ROM, NZS, and EZS (p < 0.01). The BPSS increased the rigidity (ROM, NZS, and EZS) of the intact model in lateral bending and axial rotation (p ≤ 0.01), except in EZS for left rotation (p = 0.23, Friedman test). The incorporation of posterior cages marginally increased (p = 0.05) the EZS of the BPSS construct in flexion but these interbody devices provided significant stability to the ISS construct in lateral bending and axial rotation for ROM (p = 0.02), in lateral bending for NZS (p = 0.02), and in flexion/axial rotation for EZS (p ≤ 0.03); however, both PLIF constructs demonstrated equivalent ROM and stiffness (p ≥ 0.16), except in lateral bending where the PLIF-BPSS was more stable (p = 0.02). In terms of EL, the injury increased EL in flexion-extension (p = 0.02), the ISS increased EL for lateral bending and axial rotation (p ≤ 0.03), and the BPSS decreased EL in lateral bending (p = 0.02), with respect to the intact condition. The PLIF-ISS decreased lateral bending EL with respect to the ISS condition (p = 0.02), but not enough to be smaller or, at least, equivalent, to that of the PLIF-BPSS construct (p = 0.02).

Conclusions: The ISS may be a suitable device to provide immediate flexion-extension balance after a unilateral laminotomy, but the BPSS provides greater immediate stability in lateral bending and axial rotation motions. Both PLIF constructs performed equivalently in flexion-extension and axial rotation, but the PLIF-BPSS construct is more resistant to lateral bending motions. Further biomechanical and clinical evidence is required to strongly support the recommendation of a stand-alone interspinous fusion device or as supplemental fixation to expandable posterior interbody cages.

Download full-text PDF

Source
http://dx.doi.org/10.3171/2013.10.SPINE13612DOI Listing

Publication Analysis

Top Keywords

lateral bending
40
axial rotation
24
bending axial
16
interspinous fusion
12
fusion device
12
supplemental fixation
12
interbody cages
12
nzs ezs
12
bending
11
lateral
10

Similar Publications

A hydrodynamic antenna: novel lateral line system in the tail of myliobatid stingrays.

Proc Biol Sci

January 2025

Department of Organismic and Evolutionary Biology, Harvard University, 26 Oxford Street, Cambridge, MA 02138, USA.

Eagle rays, cownose rays and manta rays (order Myliobatiformes) have a slender tail that can be longer than the animal's body length, but its function and structure are unknown. Using histology, immunohistochemistry and three-dimensional imaging with micro-computed tomography scans, we describe the anatomy and function of the tail in , the cownose ray. The tail is an extension of the vertebral column with unique morphological specializations.

View Article and Find Full Text PDF

Background: Unilateral sacral fractures with posterior ring instability represent a prevalent type of posterior pelvic ring fracture. While lumbo-pelvic fixation is recognized as a highly stable method, the sufficiency of unilateral lumbo-pelvic fixation (ULF) for such fractures remains under debate.

Purpose: This study aims to assess the biomechanical stability of ULF compared to traditional bilateral lumbo-pelvic fixation (BLF) and triangular osteosynthesis (TO), incorporating clinical observations, and previous biomechanical data.

View Article and Find Full Text PDF

Background: One possible factor contributing to low back pain is hip range of motion (ROM). However, there is inconstancy in published studies, which may be due to the heterogeneity of patients. Therefore, the present study focuses on the relationship between hip ROM and lower lumbar disc herniation (LLDH).

View Article and Find Full Text PDF

A Roadmap of Responses to Asymmetry Stress in Lipid Membranes.

J Phys Chem B

January 2025

Institute of Pharmaceutical Sciences, University of Freiburg, Freiburg 79104, Germany.

The selective insertion of membrane-impermeant amphiphiles such as detergents, (lipo)peptides, drugs, etc. into the leaflet of a membrane causes an imbalance between the intrinsic areas of the and leaflet, referred to as asymmetry stress or differential stress. The literature provides individual mechanisms of how membranes respond to such stress, which are relevant to membrane remodeling processes and leakage phenomena.

View Article and Find Full Text PDF

Stored elastic bending tension as a mediator of embryonic body folding.

Cell Rep

January 2025

Department of Genetics and Developmental Biology, Rappaport Faculty of Medicine, Technion-Israel Institute of Technology, Haifa 31096, Israel. Electronic address:

During development, amniote vertebrate embryos transform from a flat sheet into a three-dimensional cylindrical form through ventral folding of the lateral sides of the sheet (the lateral plate [LP]) and their fusion in the ventral midline. Using a chick embryo slice system, we find that the flat stage is actually a poised balance of opposing dorsal and ventral elastic bending tensions. An intact extracellular matrix (ECM) is required for generating tension, as localized digestion of ECM dissipates tension, while removal of endoderm or ectoderm layers has no significant effect.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!