Background: Besides the great importance of the issue in terms of public health, there is a lack of studies evaluating the performance of several of the currently used point of care tests (POCTs) for the detection of anti-HCV.

Objectives: To investigate the performance of two POCTs for anti-HCV detection and to assess the impact of the reading time on diagnostic performance.

Patients And Methods: A total of 307 subjects were divided into three groups (1- HCV infected; 2- other chronic liver diseases; and 3- controls). The POCTs HCV Rapid Test Bioeasy® and Imuno-Rapido HCV® were read at 3, 5, 10, 15, 20 and 30 minutes. The sensitivity and specificity of the POCTs were calculated in relation to anti-HCV detection by chemiluminescence.

Results: Valid results were obtained for all tests. When compared to the chemiluminescence, both tests showed sensitivity of 97.1% and specificity of 100%. No changes in the sensitivity or specificity of the tests were observed at different reading times and when patients with other chronic liver diseases were evaluated as a control group.

Conclusions: The POCTs evaluated in this study showed high sensitivity and specificity, with no change in the performance after the third minute of reading.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3830521PMC
http://dx.doi.org/10.5812/hepatmon.12274DOI Listing

Publication Analysis

Top Keywords

anti-hcv detection
12
sensitivity specificity
12
chronic liver
8
liver diseases
8
tests
5
pocts
5
diagnostic performance
4
performance point-of-care
4
point-of-care tests
4
tests anti-hcv
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!