A multi-well fluid loading (MFL) system was developed to deliver oscillatory subphysiologic to supraphysiologic fluid shear stresses to cell monolayers in vitro using standard multi-well culture plates. Computational fluid dynamics modeling with fluid-structure interactions was used to quantify the squeeze film fluid flow between an axially displaced piston and the well plate surface. Adjusting the cone angle of the piston base modulated the fluid pressure, velocity, and shear stress magnitudes. Modeling results showed that there was near uniform fluid shear stress across the well with a linear drop in pressure across the radius of the well. Using the MFL system, RAW 264.7 osteoclastic cells were exposed to oscillatory fluid shear stresses of 0, 0.5, 1.5, 4, 6, and 17 Pa. Cells were loaded 1 h per day at 1 Hz for two days. Compared to sub-physiologic and physiologic levels, supraphysiologic oscillatory fluid shear induced upregulation of osteoclastic activity as measured by tartrate-resistant acid phosphatase activity and formation of mineral resorption pits. Cell number remained constant across all treatment groups.