To improve cardiovascular research, there is a growing need for arterial characterization in small animals. We developed a method, ARTIC (arterial characterization) for measuring lumen diameter, distension and intima media thickness (IMT). In this study ARTIC was used to automatically characterize the aorta of premature rabbit pups. Automatic measurements were compared with manual measurements, both performed by three observers. Diameter was 769 ± 140 μm (manual) and 766 ± 142 μm (automatic), distension was 35 ± 15 μm (manual) and 40 ± 12 μm (automatic) and IMT was 84 ± 11 μm (manual) and 88 ± 8 μm (automatic) (mean ± standard deviation). The variation in the measured diameter, distension and IMT ranged from 1.1% to 26.0% (manual) and from 1.0% to 9.0% (automatic). The intra-class correlation coefficient ranged from 33.0% to 99.3% (manual) and from 76.9% to 99.6% (automatic). The evaluation revealed that it is feasible to use ARTIC on B-mode images of arteries with small dimensions, which makes it a useful tool for arterial characterization in small animals.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ultrasmedbio.2013.09.017DOI Listing

Publication Analysis

Top Keywords

diameter distension
12
arterial characterization
12
automatic measurements
8
distension intima
8
intima media
8
media thickness
8
aorta premature
8
premature rabbit
8
rabbit pups
8
b-mode images
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!