Aim: The aim of this study was to check the feasibility and accuracy of measuring antetorsion during surgery using a mobile image intensifier (IF) with computed tomography (CT) function (ISO-C 3D; Siemens, Erlangen, Germany) in comparison to a conventional multi-slice CT scanner (LightSpeed QX/I CT; GE Healthcare, VA, USA).

Materials And Methods: A total of 10 intact femora with intact soft tissue of five fresh frozen cadavers were used. After fixation on a surgical table, IF CT scans of the hip and knee were performed at both 190° and 120° of scanning rotation. Afterwards, a conventional CT scan was performed. Antetorsion was calculated according to the method of Jend et al. Analysis of variance (ANOVA) and Lin's concordance correlation coefficient (LCC) were used to test the agreement between the three measurement techniques.

Results: There was no significant difference in femoral antetorsion angle measurements between the different techniques (P>0.05). The mean time required to perform a scan using the ISO-C 3D was 9±3 min. The mean time required to measure antetorsion was 8±2 min. We found a high positive correlation between CT-based measurements and measurements performed using both the ISO-C 3D at 190° (LCC=0.99; mean difference=0.02°±1.8°) and the ISO-C 3D at 120° (LCC=0.99; mean difference=0.6°±1.5°), and a high positive correlation was also seen between both ISO-C 3D methods (LCC=0.99; mean difference=0.6°±1.7°).

Conclusions: Measuring femoral antetorsion using an intra-operative IF with CT function is a feasible and accurate method. This technique could be used when there is doubt about the antetorsion angle in the operated femur and it could help decrease the need for a separate revision surgery.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.injury.2013.10.034DOI Listing

Publication Analysis

Top Keywords

femoral antetorsion
12
antetorsion angle
8
time required
8
high positive
8
positive correlation
8
antetorsion
7
iso-c
6
intra-operative assessment
4
assessment femoral
4
antetorsion iso-c
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!