When using an electronic portal imaging device (EPID) for dosimetric verifications, the calibration of the sensitive area is of paramount importance. Two calibration methods are generally adopted: one, empirical, based on an external reference dosimeter or on multiple narrow beam irradiations, and one based on the EPID response simulation. In this paper we present an alternative approach based on an intercalibration procedure, independent from external dosimeters and from simulations, and is quick and easy to perform. Each element of a detector matrix is characterized by a different gain; the aim of the calibration procedure is to relate the gain of each element to a reference one. The method that we used to compute the relative gains is based on recursive acquisitions with the EPID placed in different positions, assuming a constant fluence of the beam for subsequent deliveries. By applying an established procedure and analysis algorithm, the EPID calibration was repeated in several working conditions. Data show that both the photons energy and the presence of a medium between the source and the detector affect the calibration coefficients less than 1%. The calibration coefficients were then applied to the acquired images, comparing the EPID dose images with films. Measurements were performed with open field, placing the film at the level of the EPID. The standard deviation of the distribution of the point-to-point difference is 0.6%. An approach of this type for the EPID calibration has many advantages with respect to the standard methods - it does not need an external dosimeter, it is not related to the irradiation techniques, and it is easy to implement in the clinical practice. Moreover, it can be applied in case of transit or nontransit dosimetry, solving the problem of the EPID calibration independently from the dose reconstruction method.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5714628 | PMC |
http://dx.doi.org/10.1120/jacmp.v14i6.4420 | DOI Listing |
J Appl Clin Med Phys
January 2025
Department of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba, Canada.
Purpose: Recently, dosimetri applications of the electronic portal imaging device (EPID) in radiotherapy have gained popularity. Confidence in the robust and reliable dosimetric performance of EPID detectors is essential for their clinical use. This study aimed to evaluate the dosimetric performance of the a-Si 1200 EPID and assess the long-term stability of its response.
View Article and Find Full Text PDFPhys Med
October 2024
Department of Medicine, Yonsei University College of Medicine, Seoul, Republic of Korea; Medical Physics and Biomedical Engineering Lab (MPBEL), Yonsei University College of Medicine, Seoul, Republic of Korea; Department of Radiation Oncology, Yonsei Cancer Center, Heavy Ion Therapy Research Institute, Yonsei University College of Medicine, Seoul, Republic of Korea.
J Appl Clin Med Phys
August 2024
Department of Radiation Physics, Zhejiang Cancer Hospital, Hangzhou Institute of Medicine (HIM), Chinese Academy of Sciences, Hangzhou, Zhejiang, PR China.
The purpose of this study is to develop an electronic portal imaging device-based multi-leaf collimator calibration procedure using log files. Picket fence fields with 2-14 mm nominal strip widths were performed and normalized by open field. Normalized pixel intensity profiles along the direction of leaf motion for each leaf pair were taken.
View Article and Find Full Text PDFBiomed Phys Eng Express
July 2024
Department of Radiation Oncology, The University of Oklahoma College of Medicine, OK, United States of America.
. Small-field measurement poses challenges. Although many high-resolution detectors are commercially available, the EPID for small-field dosimetry remains underexplored.
View Article and Find Full Text PDFJ Appl Clin Med Phys
August 2024
Department of Radiation Oncology, The Fourth Affiliated Hospital of Soochow University, Suzhou, China.
Background: It has been observed that under the single isocenter conditions, the potential shifts of the electronic portal imaging devices (EPID) may be introduced when executing portal dosimetry (PD) plans for bilateral breast cancer, pleural mesothelioma, and lymphoma. These shifts are relative to the calibration positions of EPID and result in significant discrepancies in the plan verification results.
Purpose: To explore methods including correction model and specific correction matrices to revise the data obtained from displaced EPID.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!