A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 143

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 143
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 209
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3098
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 574
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 488
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

A PHP Error was encountered

Severity: Warning

Message: Attempt to read property "Count" on bool

Filename: helpers/my_audit_helper.php

Line Number: 3100

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3100
Function: _error_handler

File: /var/www/html/application/controllers/Detail.php
Line: 574
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 488
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Comparison of intraocular pressure measurements with the Reichert Pt100, the Keeler Pulsair Intellipuff portable noncontact tonometers, and Goldmann applanation tonometry. | LitMetric

Purpose: The aim of this study was to compare intraocular pressure (IOP) measurements using 2 portable tonometers, the Keeler Pulsair Intellipuff and the Reichert PT100, with Goldmann applanation tonometry (GAT).

Materials And Methods: Prospective cross-sectional study was conducted on normotensive and hypertensive patients recruited from the outpatient clinic of the University Hospital of Grenoble. IOP was measured using 2 portable noncontact tonometers (NCTs) and GAT in one eye of each patient in a random order. Central corneal thickness was measured with the Accutome PachPen Pachymeter. The Wilcoxon and t tests were used to compare the differences between tonometers, the Pearson and the Spearman correlation tests to evaluate the correlation among the methods, and the Bland-Altman plots to evaluate the agreement among the methods. Intraclass correlation coefficient (ICC) was used to evaluate the agreement among the methods in normotensive and hypertensive patients separately.

Results: A total of 137 eyes of 137 patients were included, 104 normotensive and 33 hypertensive eyes. We found an excellent agreement between PT100 and GAT (ICC=0.77) and between Intellipuff and GAT (ICC=0.75) in normotensive patients. We found a fair to good agreement (ICC=0.67) between PT100 and GAT in hypertensive patients. The agreement we found between Intellipuff and GAT was also excellent in hypertensive patients (ICC=0.86). Both NCTs significantly overestimated IOP measurements compared with GAT in normotensive and hypertensive eyes (mean of the differences between PT100 and GAT: 1.3±2.1 and 8.1±4.6 mm Hg in normotensive and in hypertensive patients, respectively (P<0.05), and between Pulsair Intellipuff and GAT: 1.5±1.8 and 2.3±4.8 mm Hg in normotensive and in hypertensive patients, respectively (P<0.05).

Conclusions: The 2 NCTs agree well with GAT in normotensive patients. The Pulsair Intellipuff also agrees well with GAT in hypertensive patients.

Download full-text PDF

Source
http://dx.doi.org/10.1097/01.ijg.0000435776.99193.41DOI Listing

Publication Analysis

Top Keywords

normotensive hypertensive
20
hypertensive patients
20
pt100 gat
12
intraocular pressure
8
reichert pt100
8
keeler pulsair
8
pulsair intellipuff
8
portable noncontact
8
noncontact tonometers
8
goldmann applanation
8